A comparison of the In, Sn, Zn and Al fixed points by radiation thermometry between NIM and NPL and verification of the NPL blackbody reference sources from 156 °C to 1000 °C

被引:9
作者
Hao, Xiaopeng [1 ]
McEvoy, Helen [2 ]
Machin, Graham [2 ]
Yuan, Zundong [1 ]
Wang, Tiejun [1 ]
机构
[1] Natl Inst Metrol, Beijing 100013, Peoples R China
[2] Natl Phys Lab, Teddington TW11 0LW, Middx, England
关键词
temperature; radiation thermometry; fixed points; blackbody radiation; CALIBRATION;
D O I
10.1088/0957-0233/24/7/075004
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
National Institute of Metrology, China (NIM) and National Physical Laboratory, UK (NPL) blackbody cavities based on the ITS-90 freezing points of In, Sn, Zn and Al have been compared by radiation thermometry. The average differences between In, Sn and Zn, measured by NPL InGaAs radiation thermometer, are 5.3, 2.2 and -7.3 mK, respectively, and the average differences between Sn, Zn and Al, measured by NIM InGaAs radiation thermometer, are 1.4, -5.8 and 13.7 mK, respectively. These differences are well within the uncertainties of the comparison. The NPL blackbody reference sources from 156 degrees C to 1000 degrees C were verified using both the NPL and the NIM InGaAs thermometers calibrated at the fixed points. The differences between the radiance temperature of the sources and the temperature measured by the contact sensor within the source are generally within 0.1 K.
引用
收藏
页数:6
相关论文
共 15 条
[1]  
[Anonymous], 1990, INF INT TEMP SCAL S
[2]  
[Anonymous], P TEMPEMKO 1999 7 IN
[3]   THE NPL REFERENCE SOURCES OF BLACKBODY RADIATION [J].
CHU, B ;
MCEVOY, HC ;
ANDREWS, JW .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1994, 5 (01) :12-19
[4]  
Chu B, 2003, AIP CONF PROC, V684, P571
[5]  
GUTSCHWAGER B, 1999, P TEMPMEKO 99 7 INT, P567
[6]  
Hao X, 2013, 9 INT TEMP S LOS ANG
[7]   Characterizing the NRC Blackbody Sources for Radiation Thermometry from 150°C to 962°C [J].
Hill, K. D. ;
Woods, D. J. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2009, 30 (01) :105-123
[8]   High-quality blackbody sources for infrared thermometry and thermography between -40 and 1000 °C [J].
Machin, G ;
Chu, B .
IMAGING SCIENCE JOURNAL, 2000, 48 (01) :15-22
[9]  
MCEVOY HC, 1997, P TEMPMEKO 96, P245
[10]   Radiation temperature scales between the indium and silver points realized at IMGC, NPL and UME using a fixed-point calibration technique [J].
Ricolfi, T ;
Battuello, M ;
Girard, F ;
Machin, G ;
McEvoy, H ;
Ugur, S ;
Diril, A .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2002, 13 (12) :2090-2093