Inhibition of biventricular pacemakers by oversensing of far-field atrial depolarization
被引:21
作者:
Lipchenca, I
论文数: 0引用数: 0
h-index: 0
机构:Chaim Sheba Med Ctr, Inst Heart, IL-52621 Tel Hashomer, Israel
Lipchenca, I
Garrigue, S
论文数: 0引用数: 0
h-index: 0
机构:Chaim Sheba Med Ctr, Inst Heart, IL-52621 Tel Hashomer, Israel
Garrigue, S
Glikson, M
论文数: 0引用数: 0
h-index: 0
机构:
Chaim Sheba Med Ctr, Inst Heart, IL-52621 Tel Hashomer, IsraelChaim Sheba Med Ctr, Inst Heart, IL-52621 Tel Hashomer, Israel
Glikson, M
[1
]
Barold, SS
论文数: 0引用数: 0
h-index: 0
机构:Chaim Sheba Med Ctr, Inst Heart, IL-52621 Tel Hashomer, Israel
Barold, SS
Clémenty, J
论文数: 0引用数: 0
h-index: 0
机构:Chaim Sheba Med Ctr, Inst Heart, IL-52621 Tel Hashomer, Israel
Clémenty, J
机构:
[1] Chaim Sheba Med Ctr, Inst Heart, IL-52621 Tel Hashomer, Israel
[2] Tel Aviv Univ, Tel Hashomer, Israel
[3] Hosp Cardiol Haut Leveque, Bordeaux, France
[4] Broward Gen Hosp, Ft Lauderdale, FL USA
来源:
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
|
2002年
/
25卷
/
03期
关键词:
biventricular pacing;
left ventricular pacing;
congestive heart failure;
pacemaker P wave oversensing;
D O I:
10.1046/j.1460-9592.2002.00365.x
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
This report describes two patients who exhibited far field oversensing of the P wave by the ventricular channel of a DDD biventricular pacemaker implanted for the treatment of congestive heart failure. Oversensing probably occurred secondary to slight displacement of the left ventricular lead in the coronary venous system. Long-term reliable pacing was restored by decreasing the sensitivity of the ventricular channel.