Quantitative analysis of the wear and wear debris from low and high carbon content cobalt chrome alloys used in metal on metal total hip replacements

被引:128
作者
Tipper, JL
Firkins, PJ
Ingham, E
Fisher, J [1 ]
Stone, MH
Farrar, R
机构
[1] Univ Leeds, Dept Microbiol, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Dept Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[3] Leeds Gen Infirm, Dept Orthopaed Surg, Leeds LS1 3EX, W Yorkshire, England
[4] Johnson & Johnson Profess, New Milton, England
基金
英国工程与自然科学研究理事会;
关键词
D O I
10.1023/A:1026473723777
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
The biological reactions to polyethylene wear debris have been shown to result in osteolysis and loosening of total hip arthroplasties. This has led to renewed interest in the use of metal on metal bearings in hip prostheses. This study employed uniaxial and biaxial multistation pin on plate reciprocators to assess how the carbon content of the cobalt chrome alloy and the types of motion affected the wear performance of the bearing surfaces and the morphology of the wear debris generated. The low carbon specimens demonstrated higher wear factors than both the mixed carbon pairings and the high carbon pairings. The biaxial motion decreased the wear rates of all specimens. Plate wear was significantly reduced by the biaxial motion, compared to pin wear. The metal wear particles isolated were an order of magnitude smaller than polyethylene particles, at 60-90 nm, and consequently, 100-fold more particles were produced per unit volume of wear compared to polyethylene. The low carbon specimens produced significantly larger particles than the other material combinations, although it is thought unlikely that the difference would be biologically significant in vivo. The volumetric wear rates were affected by the carbon content of the cobalt chrome alloy, the material combination used and type of motion applied. However, particle morphology was not affected by the carbon content of the alloy or the type of motion applied. (C) 1999 Kluwer Academic Publishers.
引用
收藏
页码:353 / 362
页数:10
相关论文
共 42 条
[1]
METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]
AMSTUTZ HC, 1996, CLIN ORTHOP S, V329, pS297
[3]
AMSTUTZ HC, 1992, CLIN ORTHOP RELAT R, V267, P7
[4]
RESORPTION OF BONE BY INFLAMMATORY CELLS DERIVED FROM THE JOINT CAPSULE OF HIP ARTHROPLASTIES [J].
ATHANASOU, NA ;
QUINN, J ;
BULSTRODE, CJK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :57-62
[5]
BENTLEY G, 1973, CLIN ORTHOP RELAT R, V95, P127
[6]
BESONG AA, 1998, IN PRESS J BONE SURG
[7]
BESONG AA, 1998, T ORTHOP RES SOC, V44, P773
[8]
Buly R L, 1992, J Arthroplasty, V7, P315, DOI 10.1016/0883-5403(92)90056-V
[9]
Metal versus polyethylene wear particles in total hip replacements [J].
Doorn, PF ;
Campbell, PA ;
Amstutz, HC .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1996, (329) :S206-S216
[10]
Doorn PF, 1996, CLIN ORTHOP RELAT R, pS187