Scanning electron microscopic investigation of catheters for blood access

被引:7
作者
Bambauer, R [1 ]
Schiel, R [1 ]
Mestres, P [1 ]
Klinkmann, J [1 ]
Sioshansi, P [1 ]
机构
[1] UNIV SAARLAND,W-6650 HOMBURG,GERMANY
关键词
large-bore catheters; surface treatment; thrombus-resistant catheter; infection-resistant catheter;
D O I
10.1159/000170268
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Using blood-contacting catheters, infections, thromboses, and stenoses are among the most frequent complications. They are caused by surface properties of the basic material. Ion beam based processes such as ion implantation (silicone rubber) and ion beam assisted deposition (silver-based coatings) affect only the outer micron of the treated material surface; there is little effect on bulk properties. These processes were employed also in the production of large-bore catheters used for extracorporeal detoxification. In a prospective study in 56 patients, 72 large-bore catheters were inserted into the internal jugular and subclavian veins and investigated after removal for bacterial colonization using a scanning electron microscope. In 24 patients 30 surface-treated catheters (Spi-Argent I n = 14; Spi-Silicone n = 16) were used. 42 untreated catheters used in 32 patients served as controls. Bacterial colonization was observed in 8.9% in contrast to 38.1% in untreated catheters. The scanning electron microscopic investigations showed in all catheters a low thrombogenicity. The ion beam based processes make the surface-treated catheters thrombus and infection resistant.
引用
收藏
页码:249 / 256
页数:8
相关论文
共 13 条
  • [1] RASTER-ELECTRON-MICROSCOPIC INVESTIGATIONS IN LARGE-BORE CATHETERS FOR EXTRACORPOREAL DETOXIFICATION
    BAMBAUER, R
    MESTRES, P
    PIROUNG, KJ
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1990, 13 (10) : 667 - 671
  • [2] BAMBAUER R, 1982, DIALYSIS TRANSPLANT, V11, P766
  • [3] Bambauer R, 1992, ASAIO J, V38, P96, DOI 10.1097/00002480-199204000-00006
  • [4] DECUBBER A, 1978, DIALYSIS TRANSPLANT, V7, P1261
  • [5] FLYNN CT, 1980, DIALYSIS TRANSPLANT, V9, P556
  • [6] Kjellstrand C M, 1975, Clin Nephrol, V4, P37
  • [7] KRAMER P, 1977, HAMOFILTRATION, P23
  • [8] LEDERLE R M, 1979, Intensivmedizin, V16, P258
  • [9] BIOCOMPATIBILITY OF SILVER-COATED POLYURETHANE CATHETERS AND SILVER-COATED DACRON(R) MATERIAL
    OLOFFS, A
    GROSSESIESTRUP, C
    BISSON, S
    RINCK, M
    RUDOLPH, R
    GROSS, U
    [J]. BIOMATERIALS, 1994, 15 (10) : 753 - 758
  • [10] SCHWARZBECK A, 1978, T AM SOC ART INT ORG, V27, P24