Adverse Local Tissue Reaction Associated With a Modular Hip Hemiarthroplasty

被引:43
作者
Whitehouse, Michael R. [1 ]
Endo, Makoto [2 ]
Masri, Bassam A. [1 ]
机构
[1] Univ British Columbia, Dept Orthopaed, Vancouver, BC V5Z 1M9, Canada
[2] Univ British Columbia, Dept Pathol, Vancouver, BC V5Z 1M9, Canada
关键词
CORROSION; PSEUDOTUMOR; ARTHROPLASTY; PREVALENCE; RELEASE; TAPER;
D O I
10.1007/s11999-013-3133-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
The local and systemic effects of wear debris and corrosion products remain a concern in arthroplasty and reaction to corrosion or wear products from modular junctions has been reported in primary and revision total joint arthroplasties. These effects have not been reported previously for unipolar hemiarthroplasties where there is no prosthetic bearing surface to contribute to the phenomenon. This may have implications for clinical surveillance and implant design. We report the case of a 72-year-old man who had symptomatic pseudotumor formation, confirmed by pathologic examination of the excised pseudotumor, with a large-head modular hip hemiarthroplasty. Metallosis and corrosion of the modular head/neck taper junction were noted at the time of revision surgery. To our knowledge, this is the first report of pseudotumor formation where the corrosion or wear products arose from the modular junction of the implant with no bearing couple present to contribute wear debris that may influence the formation of the pseudotumor. Adverse tissue reactions to wear debris generated at prosthetic articulating surfaces and corrosion and wear products from nonarticulating prosthetic junctions have been reported. The problem has been reported to be higher in metal-on-metal bearing couples and in large-diameter hip arthroplasties. Mixed-alloy junctions appear to be more susceptible to corrosion. We believe that corrosion should be considered a possible diagnostic entity when investigating persistent symptoms after hemiarthroplasty and may be avoided with the use of monoblock components.
引用
收藏
页码:4082 / 4086
页数:5
相关论文
共 20 条
[1]
[Anonymous], J ARTHROPLASTY
[2]
ECTOPIC OSSIFICATION FOLLOWING TOTAL HIP-REPLACEMENT - INCIDENCE AND A METHOD OF CLASSIFICATION [J].
BROOKER, AF ;
BOWERMAN, JW ;
ROBINSON, RA ;
RILEY, LH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1973, A 55 (08) :1629-1632
[3]
Carli A, 2011, BULL HOSP JOINT DIS, V69, pS47
[4]
CHMELL MJ, 1995, CLIN ORTHOP RELAT R, P77
[5]
Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty [J].
Cooper, H. John ;
Della Valle, Craig J. ;
Berger, Richard A. ;
Tetreault, Matthew ;
Paprosky, Wayne G. ;
Sporer, Scott M. ;
Jacobs, Joshua J. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (18) :1655-1661
[6]
The John Charnley Award Metal-on-Metal Hip Resurfacing versus Large-diameter Head Metal-on-Metal Total Hip Arthroplasty A Randomized Clinical Trial [J].
Garbuz, Donald S. ;
Tanzer, Michael ;
Greidanus, Nelson V. ;
Masri, Bassam A. ;
Duncan, Clive P. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (02) :318-325
[7]
IN-VIVO CORROSION OF MODULAR HIP-PROSTHESIS COMPONENTS IN MIXED AND SIMILAR METAL COMBINATIONS - THE EFFECT OF CREVICE, STRESS, MOTION, AND ALLOY COUPLING [J].
GILBERT, JL ;
BUCKLEY, CA ;
JACOBS, JJ .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1993, 27 (12) :1533-1544
[8]
Corrosion at the neck-stem junction as a cause of metal ion release and pseudotumour formation [J].
Gill, I. P. S. ;
Webb, J. ;
Sloan, K. ;
Beaver, R. J. .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2012, 94B (07) :895-900
[9]
Goldberg JR, 2002, CLIN ORTHOP RELAT R, P149
[10]
JACOBS JJ, 1995, CLIN ORTHOP RELAT R, P94