Glucose microsensors based on carbon paste enzyme electrodes modified with cupric hexacyanoferrate

被引:73
作者
Wang, J [1 ]
Zhang, XJ [1 ]
Prakash, M [1 ]
机构
[1] New Mexico State Univ, Dept Chem & Biochem, Las Cruces, NM 88003 USA
基金
美国国家卫生研究院;
关键词
glucose; biosensors; cupric hexacyanoferrate; hexacyanometallate; hydrogen peroxide;
D O I
10.1016/S0003-2670(99)00306-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The incorporation of cupric hexacyanoferrate within a glucose-oxidase containing carbon paste microelectrode (200 mu m i.d.) results in a highly selective, fast-responding, reagentless glucose biosensor. The hexacyanometallate modifier retains its strong electroacatalytic action towards the reduction of the enzymatically liberated hydrogen peroxide upon confinement within the carbon-paste matrix, to allow tuning of the operating potential to a region (-0.2 to 0.0 V) where unwanted reactions are negligible. The use of cupric hexacyanoferrate circumvents stability problems inherent to Prussian-Blue films in alkaline solutions, ans hence permits convenient operation at the physiological pH of 7.4. Various parameters affecting the response have been examined and optimized. Such one-step route for preparing hexacyanometallate-modified bioelectrodes obviates the need for multistep electrodepostion/polymer-casting schemes, and holds great promise for selective detection of glucose in biological fluids. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:11 / 16
页数:6
相关论文
共 10 条
[1]  
[Anonymous], 1994, ANAL ELECTROCHEMISTR
[2]   AMPEROMETRIC BIOSENSORS BASED ON THE IMMOBILIZATION OF OXIDASES IN A PRUSSIAN BLUE FILM BY ELECTROCHEMICAL CODEPOSITION [J].
CHI, QJ ;
DONG, SJ .
ANALYTICA CHIMICA ACTA, 1995, 310 (03) :429-436
[3]   PLASTICIZED POLY(VINYL CHLORIDE) AS A PERMSELECTIVE BARRIER MEMBRANE FOR HIGH-SELECTIVITY AMPEROMETRIC SENSORS AND BIOSENSORS [J].
CHRISTIE, IM ;
TRELOAR, PH ;
VADGAMA, P .
ANALYTICA CHIMICA ACTA, 1992, 269 (01) :65-73
[4]   Electrocatalytic reactions of hydrogen peroxide at carbon paste electrodes modified by some metal hexacyanoferrates [J].
Garjonyte, R ;
Malinauskas, A .
SENSORS AND ACTUATORS B-CHEMICAL, 1998, 46 (03) :236-241
[5]   CARBON-PASTE ELECTRODES MODIFIED WITH ENZYMES, TISSUES, AND CELLS [J].
GORTON, L .
ELECTROANALYSIS, 1995, 7 (01) :23-45
[6]  
Henry C, 1998, ANAL CHEM, V70, p594A
[7]   CATALYSIS OF THE REDUCTION OF MOLECULAR-OXYGEN TO WATER AT PRUSSIAN BLUE MODIFIED ELECTRODES [J].
ITAYA, K ;
SHOJI, N ;
UCHIDA, I .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1984, 106 (12) :3423-3429
[8]   PRUSSIAN BLUE BASED FIRST-GENERATION BIOSENSOR - A SENSITIVE AMPEROMETRIC ELECTRODE FOR GLUCOSE [J].
KARYAKIN, AA ;
GITELMACHER, OV ;
KARYAKINA, EE .
ANALYTICAL CHEMISTRY, 1995, 67 (14) :2419-2423
[9]   REMARKABLY SELECTIVE METALLIZED-CARBON AMPEROMETRIC BIOSENSORS [J].
WANG, J ;
LU, F ;
ANGNES, L ;
LIU, J ;
SAKSLUND, H ;
CHEN, Q ;
PEDRERO, M ;
CHEN, L ;
HAMMERICH, O .
ANALYTICA CHIMICA ACTA, 1995, 305 (1-3) :3-7
[10]   A COMPARATIVE-STUDY OF PRUSSIAN-BLUE-MODIFIED GRAPHITE PASTE ELECTRODES AND SOLID GRAPHITE-ELECTRODES WITH MECHANICALLY IMMOBILIZED PRUSSIAN-BLUE [J].
ZAKHARCHUK, NF ;
MEYER, B ;
HENNIG, H ;
SCHOLZ, F ;
JAWORKSI, A ;
STOJEK, Z .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1995, 398 (1-2) :23-35