Comparative reject analysis in conventional film-screen and digital storage phosphor radiography

被引:35
作者
Peer, S
Peer, R
Walcher, M
Pohl, M
Jaschke, W
机构
[1] Univ Innsbruck Hosp, Dept Radiol, A-6020 Innsbruck, Austria
[2] Acad Radiol Technicians, A-6020 Innsbruck, Austria
关键词
reject analysis; digital radiography; quality control;
D O I
10.1007/s003300050911
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was comparative analysis of rejected radiographs in conventional and digital radiology under the aspects number of rejected images ages and reasons for rejection. During 2 months waste films of conventional radiography were collected in digital radiography each image-delete command at the postprocessing workstation was documented. Rejected images were analysed and assigned to four categories. The overall reject rate was 27.6 % in the conventional and 2.3 % in the digital department. Whereas in the conventional department the main reason for rejection was ''exposure'' and "others" (i.e. problems related to film handling), the main reason in the digital environment was "positioning". The high exposure tolerance of digital systems markedly reduces the amount of faulty images. This is not only economically rewarding, but may also reduce unnecessary X-ray exposure of patients due to image retake.
引用
收藏
页码:1693 / 1696
页数:4
相关论文
共 8 条
[1]   REJECT ANALYSIS - A PILOT PROGRAM FOR IMAGE QUALITY MANAGEMENT [J].
ARVANITIS, TN ;
PARIZEL, PM ;
DEGRYSE, HR ;
DESCHEPPER, AMA .
EUROPEAN JOURNAL OF RADIOLOGY, 1991, 12 (03) :171-176
[2]  
Bourne D, 1969, Radiography, V35, P257
[3]  
Brandt G A, 1987, Radiol Diagn (Berl), V28, P271
[4]  
GADEHOLT G, 1989, EUR J RADIOL, V9, P137
[5]   Analysis of reject x-ray films as quality assurance element in diagnostic radiology [J].
Lewentat, G ;
Bohndorf, K .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 166 (05) :376-381
[6]   INCIDENCE AND CAUSES OF REPEATED RADIOGRAPHIC EXAMINATIONS IN A COMMUNITY HOSPITAL [J].
MAZZAFERRO, RJ ;
BALTER, S ;
JANOWER, ML .
RADIOLOGY, 1974, 112 (01) :71-72
[7]  
STARGARDT A, 1983, MOD RONTGENPHOTOGR, V2, P26
[8]  
WATKINSON S, 1984, RADIOGRAPHY, V50, P593