Evaluation of ventilators used during transport of ICU patients -: a bench study

被引:42
作者
Zanetta, G [1 ]
Robert, D [1 ]
Guérin, C [1 ]
机构
[1] Hop Croix Rousse, Serv Reanimat Med & Assistance Resp, F-69004 Lyon, France
关键词
bench study; portable ventilators; transporting ICU patients;
D O I
10.1007/s00134-002-1242-5
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objectives: To evaluate portable ventilators. Design and settings: Bench study. Materials and methods: Five portable ventilators used for transporting ICU patients [Osiris 1, (ventilator a), Osiris 2, (ventilator b), Oxylog 1000, (ventilator c), Oxylog 2000, (ventilator d), AXR1a, (ventilator e)] and three ICU ventilators which can be used for this purpose [Horus, (ventilator f), T Bird, (ventilator g), and SV 300, (ventilator h)] were compared using a test lung regarding: 1) their capability to maintain set tidal volumes (V-T) of 300 ml, 500 ml, and 800 ml under a normal condition A [resistance (R) 5 cmH(2)O/l/s and compliance (C) 100 ml/cmH(2)O] and two abnormal conditions B (R 20-C 30) and C (R 50-C 100); 2) trapped volume (expired VT relative to inspired V-T at 0.7 s, 1 s, and 1.4 s), an estimate of the expiratory resistance of both circuit and valve; and 3) the triggering system assessed from the measurements of Deltat, DeltaP for two inspiratory efforts at a PEEP of 0 cmH(2)O and 5 cmH(2)O in ventilators b, d, f, g, and h. Flow and airway pressure were measured with an independent physiologic recording system. Results: 1) VT. For ventilators a-h, the mean +/- SD changes of a set VT of 300 ml were -2.6+/-0.2%, -9.7+/-0.2%, 0+/-0%, -6.1+/-0.2%, 1.0+/-0.3%, -2.1+/-1.7%, 0.3+/-0%, and -1.3+/-0.1% (P<0.001), respectively, during condition B relative to A. Similar results were obtained for a VT of 500 ml and 800 ml and during condition C relative to A; 2) Trapped volume. For ventilators a-h, trapped volume averaged 1+/-1 %, 20+/-0%, 30+/-0.4%, 20+/-1%, 1+/-0%, 19+/-0%, 15+/-0%, and 14+/-0% at 0.7 s (P <0.001) and 0.6+/-0%, 5+/-0%, 0.5+/-0%, 0+/-0%%, 0+/-0%, 0.6+/-0%, 0+/-0%, and 0+/-0% at 1.4 s (P=NS); and 3) the triggering system of Oxylog 2000 was poor whereas it was of good quality for Horus, T-Bird, SV 300, and Osiris 2. Conclusions: The small portable ventilators presently investigated varied between each other and were less accurate than ICU ventilators.
引用
收藏
页码:443 / 451
页数:9
相关论文
共 15 条
[1]   RESPIRATORY COMPLIANCE AND RESISTANCE IN MECHANICALLY VENTILATED PATIENTS WITH ACUTE RESPIRATORY-FAILURE [J].
BERNASCONI, M ;
PLOYSONGSANG, Y ;
GOTTFRIED, SB ;
MILICEMILI, J ;
ROSSI, A .
INTENSIVE CARE MEDICINE, 1988, 14 (05) :547-553
[2]   RESPIRATORY MECHANICS DURING THE 1ST DAY OF MECHANICAL VENTILATION IN PATIENTS WITH PULMONARY-EDEMA AND CHRONIC AIRWAY-OBSTRUCTION [J].
BROSEGHINI, C ;
BRANDOLESE, R ;
POGGI, R ;
POLESE, G ;
MANZIN, E ;
MILICEMILI, J ;
ROSSI, A .
AMERICAN REVIEW OF RESPIRATORY DISEASE, 1988, 138 (02) :355-361
[3]   Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. [J].
Brower, RG ;
Matthay, MA ;
Morris, A ;
Schoenfeld, D ;
Thompson, BT ;
Wheeler, A ;
Wiedemann, HP ;
Arroliga, AC ;
Fisher, CJ ;
Komara, JJ ;
Perez-Trepichio, P ;
Parsons, PE ;
Wolkin, R ;
Welsh, C ;
Fulkerson, WJ ;
MacIntyre, N ;
Mallatratt, L ;
Sebastian, M ;
McConnell, R ;
Wilcox, C ;
Govert, J ;
Thompson, D ;
Clemmer, T ;
Davis, R ;
Orme, J ;
Weaver, L ;
Grissom, C ;
Eskelson, M ;
Young, M ;
Gooder, V ;
McBride, K ;
Lawton, C ;
d'Hulst, J ;
Peerless, JR ;
Smith, C ;
Brownlee, J ;
Pluss, W ;
Kallet, R ;
Luce, JM ;
Gottlieb, J ;
Elmer, M ;
Girod, A ;
Park, P ;
Daniel, B ;
Gropper, M ;
Abraham, E ;
Piedalue, F ;
Glodowski, J ;
Lockrem, J ;
McIntyre, R .
NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (18) :1301-1308
[4]  
CAMPBELL R S, 1992, Respiratory Care, V37, P29
[5]   DEMAND AND CONTINUOUS-FLOW INTERMITTENT MANDATORY VENTILATION SYSTEMS [J].
CHRISTOPHER, KL ;
NEFF, TA ;
BOWMAN, JL ;
EBERLE, DJ ;
IRVIN, CG ;
GOOD, JT .
CHEST, 1985, 87 (05) :625-630
[6]  
CORCELLE P, 1999, RBM, V21, P148
[7]   INVESTIGATION OF THE SPONTANEOUS MODES OF BREATHING OF DIFFERENT VENTILATORS [J].
COX, D ;
TINLOI, SF ;
FARRIMOND, JG .
INTENSIVE CARE MEDICINE, 1988, 14 (05) :532-537
[8]  
DUREIL B, 2000, REANIM URGENCES, V9, P477
[9]   ANALYSIS OF BEHAVIOR OF THE RESPIRATORY SYSTEM IN ARDS PATIENTS - EFFECTS OF FLOW, VOLUME, AND TIME [J].
EISSA, NT ;
RANIERI, VM ;
CORBEIL, C ;
CHASSE, M ;
ROBATTO, FM ;
BRAIDY, J ;
MILICEMILI, J .
JOURNAL OF APPLIED PHYSIOLOGY, 1991, 70 (06) :2719-2729
[10]   RISK, COST, AND BENEFIT OF TRANSPORTING ICU PATIENTS FOR SPECIAL STUDIES [J].
INDECK, M ;
PETERSON, S ;
SMITH, J ;
BROTMAN, S .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1988, 28 (07) :1020-1025