Filtered Back Projection, Adaptive Statistical Iterative Reconstruction, and a Model-based Iterative Reconstruction in Abdominal CT: An Experimental Clinical Study

被引:229
作者
Deak, Zsuzsanna [1 ]
Grimm, Jochen M. [1 ]
Treitl, Marcus [1 ]
Geyer, Lucas L. [1 ]
Linsenmaier, Ulrich [1 ]
Koerner, Markus [1 ]
Reiser, Maximilian F. [1 ]
Wirth, Stefan [1 ]
机构
[1] Univ Munich, Dept Clin Radiol, D-80336 Munich, Germany
关键词
IMAGE-QUALITY; DOSE REDUCTION; ALGORITHM; OPTIMIZATION; ANGIOGRAPHY; PERFORMANCE; RISKS; MDCT;
D O I
10.1148/radiol.12112707
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: To compare objective and subjective image quality parameters of three image reconstruction algorithms of different generations at routine multidetector computed tomographic (CT) examinations of the abdomen. Materials and Methods: This institutional review board-approved study included 22 consecutive patients (mean age, 56.1 years +/- 15.8 [standard deviation]; mean weight, 79.1 kg +/- 14.8) who underwent routine CT examinations of the abdomen. A low-contrast phantom was used for objective quality control. Raw data sets were reconstructed by using filtered back projection (FPB), adaptive statistical iterative reconstruction (ASIR), and a model-based iterative reconstruction (MBIR). Radiologists used a semiquantitative scale (-3 to +3) to rate subjective image quality and artifacts, comparing both FBP and MBIR images with ASIR images. The Wilcoxon test and the intraclass correlation coefficient were used to evaluate the data. Measurements of objective noise and CT numbers of soft tissue structures were compared with analysis of variance. Results: The phantom study revealed an improved detectability of low-contrast targets for MBIR compared with ASIR or FBP. Subjective ratings showed higher image quality for MBIR, with better resolution (median value, 2; range, 1 to 3), lower noise (2; range, 1 to 3), and finer contours (2; range, 1 to 2) compared with ASIR (all P,.001). FBP performed inferiorly (0, range, 22 to 0]; -1 [range, -3 to 0]; 0 [range, 21 to 0], respectively; all, P,.001). Mean interobserver correlation was 0.9 for image perception and 0.7 for artifacts. Objective noise for FBP was 14%-68% higher and for MBIR was 18%-47% lower than that for ASIR (P < .001). Conclusion: The MBIR algorithm considerably improved objective and subjective image quality parameters of routine abdominal multidetector CT images compared with those of ASIR and FBP. (C) RSNA, 2012
引用
收藏
页码:197 / 206
页数:10
相关论文
共 31 条
[1]
Artifacts in CT: Recognition and avoidance [J].
Barrett, JF ;
Keat, N .
RADIOGRAPHICS, 2004, 24 (06) :1679-1691
[2]
Bongartz G, 16252 EUR
[3]
Estimated radiation risks potentially associated with full-body CT screening [J].
Brenner, DJ ;
Elliston, CD .
RADIOLOGY, 2004, 232 (03) :735-738
[4]
Bayesian statistical reconstruction for low-dose X-ray computed tomography using an adaptive-weighting nonlocal prior [J].
Chen, Yang ;
Gao, Dazhi ;
Nie, Cong ;
Luo, Limin ;
Chen, Wufan ;
Yin, Xindao ;
Lin, Yazhong .
COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2009, 33 (07) :495-500
[5]
Computed tomography-old ideas and new technology [J].
Fleischmann, Dominik ;
Boas, F. Edward .
EUROPEAN RADIOLOGY, 2011, 21 (03) :510-517
[6]
Combination of a Low-Tube-Voltage Technique With Hybrid Iterative Reconstruction (iDose) Algorithm at Coronary Computed Tomographic Angiography [J].
Funama, Yoshinori ;
Taguchi, Katsuyuki ;
Utsunomiya, Daisuke ;
Oda, Seitaro ;
Yanaga, Yumi ;
Yamashita, Yasuyuki ;
Awai, Kazuo .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2011, 35 (04) :480-485
[7]
CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT [J].
Gervaise, Alban ;
Osemont, Benoit ;
Lecocq, Sophie ;
Noel, Alain ;
Micard, Emilien ;
Felblinger, Jacques ;
Blum, Alain .
EUROPEAN RADIOLOGY, 2012, 22 (02) :295-301
[8]
CT iterative reconstruction in image space: A phantom study [J].
Ghetti, C. ;
Ortenzia, O. ;
Serreli, G. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :161-165
[9]
State of the art: Technologies for computed tomography dose reduction [J].
Gunn M.L.D. ;
Kohr J.R. .
Emergency Radiology, 2010, 17 (3) :209-218
[10]
Cancer risks from diagnostic radiology [J].
Hall, E. J. ;
Brenner, D. J. .
BRITISH JOURNAL OF RADIOLOGY, 2008, 81 (965) :362-378