Design, assembly, and optical bench testing of a high-numerical-aperture miniature injection-molded objective for fiber-optic confocal reflectance microscopy

被引:29
作者
Chidley, MD [1 ]
Carlson, KD
Richards-Kortum, RR
Descour, MR
机构
[1] Univ Arizona, Coll Opt Sci, Tucson, AZ 85721 USA
[2] Univ Texas, Dept Biomed Engn, Austin, TX 78712 USA
[3] Rice Univ, Dept Bioengn, Houston, TX 77005 USA
关键词
D O I
10.1364/AO.45.002545
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The design, analysis, assembly methods, and optical-bench test results for a miniature injection-molded plastic objective lens used in a fiber-optic confocal reflectance microscope are presented. The five-lens plastic objective was tested as a stand-alone optical system before its integration into a confocal microscope for in vivo imaging of cells and tissue. Changing the spacing and rotation of the individual optical elements can compensate for fabrication inaccuracies and improve performance. The system performance of the miniature objective lens is measured by use of an industry-accepted slanted-edge modulation transfer function (MTF) metric. An estimated Strehl ratio of 0.61 and a MTF value of 0.66 at the fiber-optic bundle Nyquist frequency have been obtained. The optical bench testing system is configured to permit interactive optical alignment during testing to optimize performance. These results are part of an effort to demonstrate the manufacturability of low-cost, high-performance biomedical optics for high-resolution in vivo imaging. Disposable endoscopic microscope objectives could help in vivo confocal microscopy technology mature to permit wide-scale clinical screening and detection of early cancers and precancerous lesions. (c) 2006 Optical Society of America.
引用
收藏
页码:2545 / 2554
页数:10
相关论文
共 26 条
[1]  
BAUMER S, 2005, HDB PLASTIC OPTICS, P28
[2]  
BOREMAN GD, 2001, MODULATION TRANSFER, P28
[3]  
Burns PA, 2000, PICS 2000: IMAGE PROCESSING, IMAGE QUALITY, IMAGE CAPTURE, SYSTEMS CONFERENCE, PROCEEDINGS, P135
[4]  
Burns PD, 2002, PICS 2002: IMAGE PROCESSING, IMAGE QUALITY, IMAGE CAPTURE, SYSTEMS CONFERENCE, PROCEEDINGS, P193
[5]   In vivo fiber-optic confocal reflectance microscope with an injection-molded plastic miniature objective lens [J].
Carlson, K ;
Chidley, M ;
Sung, KB ;
Descour, M ;
Gillenwater, A ;
Follen, M ;
Richards-Kortum, R .
APPLIED OPTICS, 2005, 44 (10) :1792-1797
[6]   Miniature injection-molded optics for fiber-optic, in vivo confocal microscopy [J].
Chidley, MD ;
Liang, C ;
Descour, MR ;
Sung, KB ;
Richards-Kortum, R ;
Gillenwater, AM .
INTERNATIONAL OPTICAL DESIGN CONFERENCE 2002, 2002, 4832 :126-136
[7]   Near real-time confocal microscopy of amelanotic tissue: Detection of dysplasia in ex vivo cervical tissue [J].
Collier, T ;
Lacy, A ;
Richards-Kortum, R ;
Malpica, A ;
Follen, M .
ACADEMIC RADIOLOGY, 2002, 9 (05) :504-512
[8]   Sources of contrast in confocal reflectance imaging [J].
Dunn, AK ;
Smithpeter, C ;
Welch, AJ ;
RichardsKortum, R .
APPLIED OPTICS, 1996, 35 (19) :3441-3446
[9]  
FISHER RE, 2000, OPTICAL SYSTEM DESIG, P110
[10]   CONFOCAL MICROSCOPY THROUGH A FIBEROPTIC IMAGING BUNDLE [J].
GMITRO, AF ;
AZIZ, D .