摘要
For in vivo magnetic resonance imaging at high field (>= 3 T) it is essential to consider the homogeneity of the active B, field (B-1+), particularly if surface coils are used for RF transmission. A new method is presented for highly rapid B-1+ magnitude mapping. It combines the double angle method with a B-1-insensitive magnetization-reset sequence such that the choice of repetition time (TR) is independent of T, and with a multislice segmented (spiral) acquisition to achieve volumetric coverage with adequate spatial resolution in a few seconds. Phantom experiments confirmed the accuracy of this technique even when TR << T-1, with the side effect being lowered SNR. The speed of this method enabled B-1+ mapping in the chest and abdomen within a single breath-hold. In human cardiac imaging, the method enabled whole-heart coverage within a single 16-s breath-hold. Results from phantoms and healthy volunteers at 1.5 T and 3 T are presented.
引用
收藏
页码:1326 / 1333
页数:8
相关论文
共 29 条
[1]   HIGH-SPEED SPIRAL-SCAN ECHO PLANAR NMR IMAGING .1. [J].
AHN, CB ;
KIM, JH ;
CHO, ZH .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1986, 5 (01) :2-7
[2]   RADIOFREQUENCY MAP OF AN NMR COIL BY IMAGING [J].
AKOKA, S ;
FRANCONI, F ;
SEGUIN, F ;
LEPAPE, A .
MAGNETIC RESONANCE IMAGING, 1993, 11 (03) :437-441
[3]  
Bracewell R. N., 1986, FOURIER TRANSFORM IT, V31999
[4]   Different excitation and reception distributions with a single-loop transmit-receive surface coil near a head-sized spherical phantom at 300 MHz [J].
Collins, CM ;
Yang, QX ;
Wang, JH ;
Zhang, X ;
Liu, H ;
Michaeli, S ;
Zhu, XH ;
Adriany, G ;
Vaughan, JT ;
Anderson, P ;
Merkle, H ;
Ugurbil, K ;
Smith, MB ;
Chen, W .
MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (05) :1026-1028
[5]   Optimized RF excitation for anatomical brain imaging of the occipital lobe using the 3D MDEFT sequence and a surface transmit coil [J].
Deichmann, R .
MAGNETIC RESONANCE IN MEDICINE, 2005, 53 (05) :1212-1216
[6]   Double inversion black-blood fast spin-echo imaging of the human heart: A comparison between 1.5T and 3.0T [J].
Greenman, RL ;
Shirosky, JE ;
Mulkern, RV ;
Rofsky, NM .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2003, 17 (06) :648-655
[7]   ELECTRICALLY CALIBRATED PLETHYSMOGRAPH FOR DIRECT MEASUREMENT OF LIMB BLOOD-FLOW [J].
HOKANSON, DE ;
SUMNER, DS ;
STRANDNESS, DE .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1975, BM22 (01) :25-29
[8]   MAGNETIC-FIELD MAPPING [J].
HORNAK, JP ;
SZUMOWSKI, J ;
BRYANT, RG .
MAGNETIC RESONANCE IN MEDICINE, 1988, 6 (02) :158-163
[9]  
Hoult DI, 2000, J MAGN RESON IMAGING, V12, P46, DOI 10.1002/1522-2586(200007)12:1<46::AID-JMRI6>3.0.CO
[10]  
2-D