Quantitative comparison of proteomic data quality between a 2D and 3D quadrupole ion trap

被引:35
作者
Blackler, AR
Klammer, AA
MacCoss, MJ
Wu, CC [1 ]
机构
[1] Univ Colorado, Hlth Sci Ctr, Dept Pharmacol, Aurora, CO 80045 USA
[2] Univ Washington, Dept Genome Sci, Seattle, WA 98198 USA
关键词
D O I
10.1021/ac051486a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A 2D ion trap has a greater ion trapping efficiency, greater ion capacity before observing space-charging effects, and a faster ion ejection rate than a traditional 3D ion trap mass spectrometer. These hardware improvements should result in a significant increase in protein identifications from complex mixtures analyzed using shotgun proteomics. In this study, we compare the quality and quantity of peptide identifications using data-dependent acquisition of tandem mass spectra of peptides between two commercially available ion trap mass spectrometers (an LTQ and an LCQ XP Max). We demonstrate that the increased trapping efficiency, increased ion capacity, and faster ion ejection rate of the LTQ results in greater than 5-fold more protein identifications, better identification of low-abundance proteins, and higher confidence protein identifications when compared with a LCQ XP Max.
引用
收藏
页码:1337 / 1344
页数:8
相关论文
共 17 条
[1]   Genome-scale analysis of the uses of the Escherichia coli genome:: Model-driven analysis of heterogeneous data sets [J].
Allen, TE ;
Herrgård, MJ ;
Liu, MZ ;
Qiu, Y ;
Glasner, JD ;
Blattner, FR ;
Palsson, BO .
JOURNAL OF BACTERIOLOGY, 2003, 185 (21) :6392-6399
[2]   A new algorithm for the evaluation of shotgun peptide sequencing in proteomics: Support vector machine classification of peptide MS/MS spectra and SEQUEST scores [J].
Anderson, DC ;
Li, WQ ;
Payan, DG ;
Noble, WS .
JOURNAL OF PROTEOME RESEARCH, 2003, 2 (02) :137-146
[3]   Correlation of codon bias measures with mRNA levels:: analysis of transcriptome data from Escherichia coli [J].
Goetz, RM ;
Fuglsang, A .
BIOCHEMICAL AND BIOPHYSICAL RESEARCH COMMUNICATIONS, 2005, 327 (01) :4-7
[4]   Quantitative profiling of differentiation-induced microsomal proteins using isotope-coded affinity tags and mass spectrometry [J].
Han, DK ;
Eng, J ;
Zhou, HL ;
Aebersold, R .
NATURE BIOTECHNOLOGY, 2001, 19 (10) :946-951
[5]   Empirical statistical model to estimate the accuracy of peptide identifications made by MS/MS and database search [J].
Keller, A ;
Nesvizhskii, AI ;
Kolker, E ;
Aebersold, R .
ANALYTICAL CHEMISTRY, 2002, 74 (20) :5383-5392
[6]   Direct analysis of protein complexes using mass spectrometry [J].
Link, AJ ;
Eng, J ;
Schieltz, DM ;
Carmack, E ;
Mize, GJ ;
Morris, DR ;
Garvik, BM ;
Yates, JR .
NATURE BIOTECHNOLOGY, 1999, 17 (07) :676-682
[7]   A model for random sampling and estimation of relative protein abundance in shotgun proteomics [J].
Liu, HB ;
Sadygov, RG ;
Yates, JR .
ANALYTICAL CHEMISTRY, 2004, 76 (14) :4193-4201
[8]   Probability-based validation of protein identifications using a modified SEQUEST algorithm [J].
MacCoss, MJ ;
Wu, CC ;
Yates, JR .
ANALYTICAL CHEMISTRY, 2002, 74 (21) :5593-5599
[9]   Evaluation and optimization of ion-current ratio measurements by selected-ion-monitoring mass spectrometry [J].
MacCoss, MJ ;
Toth, MJ ;
Matthews, DE .
ANALYTICAL CHEMISTRY, 2001, 73 (13) :2976-2984
[10]   Systematic comparison of a two-dimensional ion trap and a three-dimensional ion trap mass spectrometer in proteomics [J].
Mayya, V ;
Rezaul, K ;
Cong, YS ;
Han, D .
MOLECULAR & CELLULAR PROTEOMICS, 2005, 4 (02) :214-223