Thoracic electrical impedance tomographic measurements during volume controlled ventilation-effects of tidal volume and positive end-expiratory pressure

被引:92
作者
Frerichs, I [1 ]
Hahn, G [1 ]
Hellige, G [1 ]
机构
[1] Univ Gottingen, Dept Anesthesiol Res, Ctr Anesthesiol Emergency & Intens Care Med, D-37075 Gottingen, Germany
关键词
MT; PEEP; thoracic impedance;
D O I
10.1109/42.802754
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The aim of the study was to analyze thoracic electrical impedance tomographic (EIT) measurements accomplished under conditions comparable with clinical situations during artificial ventilation. Multiple EIT measurements were performed in pigs in three transverse thoracic planes during the volume controlled mode of mechanical ventilation at various tidal volumes (V-T) and positive end-expiratory pressures (PEEP), The protocol comprised following ventilatory patterns: 1) V-T (400, 500, 600, 700 mi) was varied ia a random order at various constant PEEP levels and 2) PEEP (2, 5, 8, 11, 14 cm H2O) was randomly modified during ventilation with a constant V-T. The EIT technique was used to generate cross-sectional images of 1) regional lung ventilation and 2) regional shifts in lung volume with PEEP. The quantitative analysis was performed in terms of the tidal amplitude of the impedance change, reflecting the volume of delivered gas at various preset V-T and the end-expiratory impedance change, revealing the variation of the lung volume at various PEEP levels. The results showed: 1) an increase in the tidal amplitude of the impedance change, proportional to the delivered V-T at all constant PEEP levels, 2) a rising end-expiratory impedance change, with PEEP reflecting an increase in gas volume, and 3) a PEEP-dependent redistribution of the ventilated gas between the planes. The generated images and the quantitative results indicate the ability of EIT to identify regional changes in V-T and lung volume during mechanical ventilation.
引用
收藏
页码:764 / 773
页数:10
相关论文
共 29 条
[1]   Electrical impedance tomography can monitor dynamic hyperinflation in dogs [J].
Adler, A ;
Shinozuka, N ;
Berthiaume, Y ;
Guardo, R ;
Bates, JHT .
JOURNAL OF APPLIED PHYSIOLOGY, 1998, 84 (02) :726-732
[2]   Monitoring changes in lung air and liquid volumes with electrical impedance tomography [J].
Adler, A ;
Amyot, R ;
Guardo, R ;
Bates, JHT ;
Berthiaume, Y .
JOURNAL OF APPLIED PHYSIOLOGY, 1997, 83 (05) :1762-1767
[3]   QUANTIFICATION IN IMPEDANCE IMAGING [J].
BARBER, DC .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1990, 11 :45-56
[4]  
Brown B H, 1987, Clin Phys Physiol Meas, V8 Suppl A, P91, DOI 10.1088/0143-0815/8/4A/012
[5]   EITS changes following oleic acid induced lung water [J].
Brown, BH ;
Flewelling, R ;
Griffiths, H ;
Harris, ND ;
Leathard, AD ;
Lu, L ;
Morice, AH ;
Neufeld, GR ;
Nopp, P ;
Wang, W .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 :A117-A130
[6]  
BROWN BH, 1987, MED PROG TECHNOL, V13, P69
[7]   CLINICAL-APPLICATIONS OF ELECTRICAL-IMPEDANCE TOMOGRAPHY IN THE MONITORING OF CHANGES IN INTRATHORACIC FLUID VOLUMES [J].
CAMPBELL, JH ;
HARRIS, ND ;
ZHANG, F ;
BROWN, BH ;
MORICE, AH .
PHYSIOLOGICAL MEASUREMENT, 1994, 15 :A217-A222
[8]   A parallel broadband real-time system for electrical impedance tomography [J].
Casas, O ;
Rosell, J ;
Bragos, R ;
Lozano, A ;
Riu, PJ .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 :A1-A6
[9]   Gravity-dependent phenomena in lung ventilation determined by functional EIT [J].
Frerichs, I ;
Hahn, G ;
Hellige, G .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 :A149-A157
[10]   Heterogeneous distribution of pulmonary ventilation in intensive care patients detected by functional electrical impedance tomography [J].
Frerichs, I ;
Golisch, W ;
Hahn, G ;
Kurpitz, M ;
Burchardi, H ;
Hellige, G .
JOURNAL OF INTENSIVE CARE MEDICINE, 1998, 13 (04) :168-173