Adverse tissue reactions to wear particles form Co-alloy articulations, increased by alumina-blasting particle contamination from cementless Ti-based total hip implants -: A report of seven revisions with early failure

被引:61
作者
Böhler, M [1 ]
Kanz, F [1 ]
Schwarz, B [1 ]
Steffan, I [1 ]
Walter, A [1 ]
Plenk, H [1 ]
Knahr, K [1 ]
机构
[1] Univ Vienna, Inst Histol & Embryol, A-1090 Vienna, Austria
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2002年 / 84B卷
关键词
D O I
10.1302/0301-620X.84B1.11324
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
W e revised seven alumina-blasted cementless hip v v prostheses (Ti-alloy stems, cp Ti threaded sockets) with low- or high-carbon Co-alloy bearings at a mean of 20.1 months after implantation because of pain and loosening. Histological examination of the retrieved periprosthetic tissues from two cases in which the implant was stable and three in which the socket was loose showed macrophages with basophilic granules containing metal and alumina wear particles and lymph-cell infiltrates. In one of the two cases of stem loosening the thickened neocapsule also contained definite lymphatic follicles and gross lymphocyte/plasma-cell infiltrates. Spectrometric determination of the concentration of elements in periprosthetic tissues from six cases was compared with that of joint capsules from five control patients undergoing primary hip surgery. In the revisions the mean concentration of implant-relevant elements was 693.85 mug/g dry tissue. In addition to Cr (15.2%), Co (4.3%), and Ti (10.3%), Al was predominant (68.1%) and all concentrations were significantly higher (p < 0.001) than those in the control tissues. The annual rates of linear wear were calculated for six implants. The mean value was 11.1 mum (heads 6.25 mum, inserts 4.82 mum). SEM/EDXA showed numerous fine scratches and deep furrows containing alumina particles in loosened sockets, and stems showed contamination with adhering or impacted alumina particles of between 2 and 50 mum in size.
引用
收藏
页码:128 / 136
页数:9
相关论文
共 22 条
[1]  
[Anonymous], OSTEOLOGIE
[2]  
BLACK J, 1996, CLIN ORTHOP S, V329, P244, DOI DOI 10.1097/00003086-199608001-00021
[3]  
BOEHLER M, 1999, EORS M BRUSS
[4]   Wear debris from two different alumina-on-alumina total hip arthroplasties [J].
Böhler, M ;
Mochida, Y ;
Bauer, TW ;
Plenk, H ;
Salzer, M .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2000, 82B (06) :901-909
[5]   Metal-on-metal bearing with THA and chronic renal failure. [J].
Brodner, V ;
Grohs, JG ;
Bitzan, P ;
Meisinger, V ;
Kovarik, J ;
Kotz, R .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2000, 138 (05) :425-429
[6]   Elevated serum cobalt with metal-on-metal articulating surfaces [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
GottsaunerWolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :316-321
[7]   CONTAMINATION OF TITANIUM CASTINGS BY ALUMINUM-OXIDE BLASTING [J].
DARVELL, BW ;
SAMMAN, N ;
LUK, WK ;
CLARK, RKF ;
TIDEMAN, H .
JOURNAL OF DENTISTRY, 1995, 23 (05) :319-322
[8]  
FISHER J, 1999, RELIABILITY LONG TER, P78
[9]  
GOLDBERG VM, 1995, CLIN ORTHOP RELAT R, P122
[10]   Hypersensitivity to metallic biomaterials: a review of leukocyte migration inhibition assays [J].
Hallab, N ;
Jacobs, JJ ;
Black, J .
BIOMATERIALS, 2000, 21 (13) :1301-1314