Sol-gel-based biosensor for use in stroke treatment

被引:21
作者
Grant, SA [1 ]
Glass, RS
机构
[1] Michigan Technol Univ, Ctr Biomed Engn, Houghton, MI 49931 USA
[2] Univ Calif Lawrence Livermore Natl Lab, Chem & Mat Sci Directorate, Livermore, CA 94550 USA
关键词
antibodies; biosensor; fiber optic; sol-gel; stroke;
D O I
10.1109/10.790497
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A fiber optic biosensor for the detection of fibrinolytic tic products produced during lysis of "soft" blood clots is described. The biosensor was constructed to be selective toward D dimer antigens, which form from the dissolution of cross-linked fibrin clots. The presence of D dimer antigens above a threshold level is a clinical diagnostic used to determine the presence of such occlusions following a stroke. Fluorescein-labeled D dimer antibodies are immobilized on the tip of an optical fiber by dip coating from a silica sol-gel solution. When D dimer antigens combine with the antibodies, fluorescence intensity decreases. The response of the sensor was examined in phosphate buffered saline, human plasma, and blood. Calibration plots for the sensor mere obtained in the clinically significant D dimer concentration range from 0.53 mu g/ml to 6 mu g/ml. Changes in spectroscopic properties as the sol-gel encapsulated tagged antibodies aged mere examined; a decrease in fluorescence intensity with age was noted. The D dimer antibodies remain viable for at least 4 weeks while encapsulated in the sol-gel network when stored at 4 degrees C in PBS solution. This novel sensor is being developed for use with other catheter-based microtools to treat stroke resulting from occlusion in the vascular system.
引用
收藏
页码:1207 / 1211
页数:5
相关论文
共 15 条
  • [1] ANDERSON GP, 1994, IEEE ENG MED BIO JUN, P358
  • [2] ELECTROCHEMICAL PROBING OF THE ACTIVITY OF GLUCOSE-OXIDASE EMBEDDED SOL-GEL MATRICES
    AUDEBERT, P
    DEMAILLE, C
    SANCHEZ, C
    [J]. CHEMISTRY OF MATERIALS, 1993, 5 (07) : 911 - 913
  • [3] BIOCHEMICALLY ACTIVE SOL-GEL GLASSES - THE TRAPPING OF ENZYMES
    BRAUN, S
    RAPPOPORT, S
    ZUSMAN, R
    AVNIR, D
    OTTOLENGHI, M
    [J]. MATERIALS LETTERS, 1990, 10 (1-2) : 1 - 5
  • [4] BIOCATALYSIS BY SOL-GEL ENTRAPPED ENZYMES
    BRAUN, S
    SHTELZER, S
    RAPPOPORT, S
    AVNIR, D
    OTTOLENGHI, M
    [J]. JOURNAL OF NON-CRYSTALLINE SOLIDS, 1992, 147 : 739 - 743
  • [5] SOL-GEL ENCAPSULATION METHODS FOR BIOSENSORS
    DAVE, BC
    DUNN, B
    VALENTINE, JS
    ZINK, JI
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (22) : A1120 - A1127
  • [6] ENCAPSULATION OF PROTEINS IN TRANSPARENT POROUS SILICATE-GLASSES PREPARED BY THE SOL-GEL METHOD
    ELLERBY, LM
    NISHIDA, CR
    NISHIDA, F
    YAMANAKA, SA
    DUNN, B
    VALENTINE, JS
    ZINK, JI
    [J]. SCIENCE, 1992, 255 (5048) : 1113 - 1115
  • [7] PLASMIC DEGRADATION OF CROSSLINKED FIBRIN - CHARACTERIZATION OF NEW MACROMOLECULAR SOLUBLE COMPLEXES AND A MODEL OF THEIR STRUCTURE
    FRANCIS, CW
    MARDER, VJ
    BARLOW, GH
    [J]. JOURNAL OF CLINICAL INVESTIGATION, 1980, 66 (05) : 1033 - 1043
  • [8] FLUOROMETER AND TAPERED FIBER OPTIC PROBES FOR SENSING IN THE EVANESCENT WAVE
    GOLDEN, JP
    SHRIVERLAKE, LC
    ANDERSON, GP
    THOMPSON, RB
    LIGLER, FS
    [J]. OPTICAL ENGINEERING, 1992, 31 (07) : 1458 - 1462
  • [9] AN EVANESCENT-WAVE BIOSENSOR .2. FLUORESCENT SIGNAL ACQUISITION FROM TAPERED FIBER OPTIC PROBES
    GOLDEN, JP
    ANDERSON, GP
    RABBANY, SY
    LIGLER, FS
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1994, 41 (06) : 585 - 591
  • [10] CLINICAL-ASSESSMENT OF STROKE
    GOLDSTEIN, LB
    MATCHAR, DB
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1994, 271 (14): : 1114 - 1120