The Camino intracranial pressure device in clinical practice.: Assessment in a 1000 cases

被引:107
作者
Gelabert-González, M
Ginesta-Galan, V
Sernamito-García, R
Allut, AG
Bandin-Diéguez, J
Rumbo, RM
机构
[1] Univ Santiago de Compostela, Hosp Santiago de Compostela, Dept Surg, Neurosurg Serv, Santiago De Compostela 15706, Spain
[2] Univ Santiago de Compostela, Hosp Santiago de Compostela, Dept Surg, Anaesthesia & Reanimat Serv, Santiago De Compostela 15706, Spain
关键词
Camino device; complications; fiberoptic device; intracranial pressure; ICP monitoring; zero drift;
D O I
10.1007/s00701-005-0683-3
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background. Intracranial pressure (ICP) monitoring has become standard in the management of neurocritical patients. A variety of monitoring techniques and devices are available, each offering advantages and disadvantages. Analysis of large populations has never been performed. Patients and methods. A prospective study was designed to evaluate the Camino (R) fiberoptic intraparenchymal cerebral pressure monitor for complications and accuracy. Results. Between 1992-2004 one thousand consecutive patients had a fiberoptic ICP monitor placed. The most frequent indication for monitoring was severe head injury (697 cases). The average duration of ICP monitoring was 184.6 +/- 94.3 hours; the range was 16-581 hours. Zero drift (range, -17 to 21 mm Hg; mean 7.3 +/- 5.1) was recorded after the devices were removed from 624 patients. Mechanical complications such as: breakage of the optical fiber (n = 17); dislocations of the fixation screw (n = 15) or the probe (n = 13); and failure of ICP recording for unknown reasons (n = 4) were found in 49 Camino (R) devices. Conclusions. The Camino ICP sensor remains one of the most popular ICP monitoring devices for use in critical neurosurgical patients. The system offers reliable ICP measurements in an acceptable percentage of device complications and the advantage of in vivo recalibration. The incidence of technical complications was low and similar to others devices.
引用
收藏
页码:435 / 441
页数:7
相关论文
共 32 条
[1]   Complications of intracranial pressure monitoring in children with head trauma [J].
Anderson, RCE ;
Kan, P ;
Klimo, P ;
Brockmeyer, DL ;
Walker, ML ;
Kestle, JRW .
JOURNAL OF NEUROSURGERY, 2004, 101 (01) :53-58
[2]   Prospective study of zero drift in fiberoptic pressure monitors used in clinical practice [J].
Bavetta, S ;
Norris, JS ;
Wyatt, M ;
Sutcliffe, JC ;
Hamlyn, PJ .
JOURNAL OF NEUROSURGERY, 1997, 86 (06) :927-930
[3]   Complications of brain tissue pressure monitoring with a fiberoptic device [J].
Bekar, A ;
Gören, S ;
Korfali, E ;
Aksoy, K ;
Boyaci, S .
NEUROSURGICAL REVIEW, 1998, 21 (04) :254-259
[4]   Traumatic brain injury and haemorrhagic complications after intracranial pressure monitoring [J].
Blaha, M ;
Lazar, D .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2005, 76 (01) :147-147
[5]  
Bullock RM, 2000, J NEUROTRAUM, V17, P449
[6]  
CHAMBERS IR, 1993, NEUROSURGERY, V33, P866
[7]   Intracranial pressure - Part two: Clinical applications and technology [J].
Citerio, G ;
Andrews, PJD .
INTENSIVE CARE MEDICINE, 2004, 30 (10) :1882-1885
[8]   Laboratory testing of three intracranial pressure microtransducers: Technical report [J].
Czosnyka, M ;
Czosnyka, Z ;
Pickard, JD .
NEUROSURGERY, 1996, 38 (01) :219-224
[9]   Laboratory testing of the Spiegelberg brain pressure monitor: a technical report [J].
Czosnyka, M ;
Czosnyka, Z ;
Pickard, JD .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1997, 63 (06) :732-735
[10]   Placement of intracranial pressure monitors: Are "normal" coagulation parameters necessary? [J].
Davis, JW ;
Davis, IC ;
Bennink, LD ;
Hysell, SE ;
Curtis, BV ;
Kaups, KL ;
Bilello, JF .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2004, 57 (06) :1173-1177