Prospective study of zero drift in fiberoptic pressure monitors used in clinical practice

被引:22
作者
Bavetta, S
Norris, JS
Wyatt, M
Sutcliffe, JC
Hamlyn, PJ
机构
[1] Department of Neurosurgery, Royal London Hospital, London
[2] Department of Neurosurgery, Royal London Hospital, Whitechapel
基金
日本学术振兴会;
关键词
fiberoptic pressure monitor; intracranial pressure measurement; intraventricular pressure; subarachnoid pressure; subdural pressure;
D O I
10.3171/jns.1997.86.6.0927
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
One hundred and one fiberoptic pressure transducers (59 subdural and 42 ventricular) were studied in 86 patients (some in whom more than one device had been inserted). Only four complications occurred: two transient cerebrospinal fluid leaks after removal and two clinically insignificant intracerebral hematomas. No intracranial infections could be attributed to the devices. Technical problems occurred 23 times, with 11 devices ceasing to function before removal, seven becoming displaced, and five microventricular catheters failing to enter the ventricles. Zero-drift readings were obtained for 83 devices at the time of removal (median 66 hours after insertion, range 2 hours-13 days). There was a clear negative bias in the readings (median -3), with a wide range of values (-12 to +14 mm Hg; interquartile range -6 to -1) that was apparent even in the first 3 days of use. There was no important relationship between zero drift and any recorded variable. It is concluded that zero drift of fiberoptic pressure transducers is a significant problem and that undue reliance should not be placed on intracranial pressure readings from these devices in isolation from other clinical and radiological information.
引用
收藏
页码:927 / 930
页数:4
相关论文
共 13 条
  • [1] TECHNICAL ASPECTS AND CLINICAL RELIABILITY OF INTRACRANIAL-PRESSURE MONITORING WITH A PARENCHYMAL FIBEROPTIC TRANSDUCER
    ARTRU, F
    TERRIER, A
    GIBERT, I
    MESSAOUDI, K
    CHARLOT, M
    NAOUS, H
    JOURDAN, C
    [J]. ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 1992, 11 (04): : 424 - 429
  • [2] A prospective comparison of fibre-optic and fluid-filled single lumen bolt subdural pressure transducers in ventilated neurosurgical patients
    Bavetta, S
    Sutcliffe, JC
    Sparrow, OCE
    Hamlyn, PJ
    [J]. BRITISH JOURNAL OF NEUROSURGERY, 1996, 10 (03) : 279 - 284
  • [3] A CLINICAL-EVALUATION OF THE CAMINO SUBDURAL SCREW AND VENTRICULAR MONITORING KITS
    CHAMBERS, IR
    MENDELOW, AD
    SINAR, EJ
    MODHA, P
    [J]. NEUROSURGERY, 1990, 26 (03) : 421 - 423
  • [4] CHAMBERS IR, 1993, NEUROSURGERY, V33, P866
  • [5] EVALUATION OF A FIBEROPTIC INTRACRANIAL-PRESSURE MONITOR
    CRUTCHFIELD, JS
    NARAYAN, RK
    ROBERTSON, CS
    MICHAEL, LH
    [J]. JOURNAL OF NEUROSURGERY, 1990, 72 (03) : 482 - 487
  • [6] EDDY VA, 1995, AM SURGEON, V61, P24
  • [7] MONITORING OF BRAIN-TISSUE PRESSURE WITH A FIBEROPTIC DEVICE
    GAMBARDELLA, G
    DAVELLA, D
    TOMASELLO, F
    HORWITZ, NH
    WILKINSON, HA
    [J]. NEUROSURGERY, 1992, 31 (05) : 918 - 922
  • [8] CONTINUOUS MONITORING OF INTRACRANIAL-PRESSURE WITH A MINIATURIZED FIBEROPTIC DEVICE
    OSTRUP, RC
    LUERSSEN, TG
    MARSHALL, LF
    ZORNOW, MH
    [J]. JOURNAL OF NEUROSURGERY, 1987, 67 (02) : 206 - 209
  • [9] INTRACRANIAL-PRESSURE MONITORING - FIBEROPTIC MONITOR COMPARED WITH THE VENTRICULAR CATHETER
    SCHICKNER, DJ
    YOUNG, RF
    [J]. SURGICAL NEUROLOGY, 1992, 37 (04): : 251 - 254
  • [10] The fiberoptic intraparenchymal cerebral pressure monitor in 244 patients
    Shapiro, S
    Bowman, R
    Callahan, J
    Wolfla, C
    [J]. SURGICAL NEUROLOGY, 1996, 45 (03): : 278 - 281