Enhanced determination of glucose by microchip electrophoresis with pulsed amperometric detection

被引:50
作者
García, CD [1 ]
Henry, CS [1 ]
机构
[1] Colorado State Univ, Dept Chem, Ft Collins, CO 80523 USA
关键词
microchip-CE; SDS; post-column modification; pulsed amperometric detection; glucose;
D O I
10.1016/j.aca.2003.11.060
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The present report describes a new analysis strategy for microchip capillary electrophoresis with pulsed amperometric detection and its application to the determination of glucose. The addition of sodium dodecyl sulfate (SDS) to the mobile phase and detection reservoir stabilized flow rates and enhanced the detection signal for glucose. A higher pH (compared to the running buffer) was used at the waste reservoir in order to improve the detection performance while maintaining good separations. To our knowledge, this is the first report describing the use of post-column pH modification using microchip electrophoresis. Under optimum conditions, a linear relationship between the peak current and the concentration of glucose was found between 10(-2)-10(-15) M, with a limit-of-detection of 1.2 muM. In addition, the separation of glucosamine and glucose was performed at pH 7.1 while the detection was performed at pH I I to demonstrate the ability to use post-column pH modification. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 73 条
  • [1] Micro total analysis systems. 2. Analytical standard operations and applications
    Auroux, PA
    Iossifidis, D
    Reyes, DR
    Manz, A
    [J]. ANALYTICAL CHEMISTRY, 2002, 74 (12) : 2637 - 2652
  • [2] Badal MY, 2002, J CHROMATOGR A, V947, P277
  • [3] Polymer microfluidic devices
    Becker, H
    Locascio, LE
    [J]. TALANTA, 2002, 56 (02) : 267 - 287
  • [4] Belder D, 2002, ELECTROPHORESIS, V23, P3567, DOI 10.1002/1522-2683(200210)23:20<3567::AID-ELPS3567>3.0.CO
  • [5] 2-3
  • [6] Chiari M, 2002, ELECTROPHORESIS, V23, P536, DOI 10.1002/1522-2683(200202)23:4<536::AID-ELPS536>3.0.CO
  • [7] 2-#
  • [8] Characterization of electrophoretic behavior of sugar isomers by microchip electrophoresis coupled with videomicroscopy
    Dang, FQ
    Zhang, LH
    Jabasini, M
    Kaji, N
    Baba, Y
    [J]. ANALYTICAL CHEMISTRY, 2003, 75 (10) : 2433 - 2439
  • [9] Enantioselective uptake of amino acids using an electromodulated column packed with carbon fibres modified with overoxidised polypyrrole
    Deore, B
    Yakabe, H
    Shiigi, H
    Nagaoka, T
    [J]. ANALYST, 2002, 127 (07) : 935 - 939
  • [10] Comparison of standard capillary and chip separations of sodium dodecylsulfate-protein complexes
    Deyl, Z
    Miksík, I
    Eckhardt, A
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2003, 990 (1-2) : 153 - 158