Effect of corneal parameters on measurements using the pulsatile ocular blood flow tonograph and Goldmann applanation tonometer

被引:86
作者
Gunvant, P
Baskaran, M
Vijaya, L
Joseph, IS
Watkins, RJ
Nallapothula, M
Broadway, DC
O'Leary, DJ
机构
[1] Anglia Polytech Univ, Dept Optometry & Ophthalm Dispensing, Cambridge, England
[2] Univ Louisville, Dept Psychol & Brain Sci, Vis Sci Grp, Louisville, KY 40292 USA
[3] Med Res Fdn, Madras 600006, Tamil Nadu, India
[4] Univ Bradford, Dept Optometry, Bradford BD7 1DP, W Yorkshire, England
[5] Norfolk & Norwich Univ Hosp NHS Trust, Dept Ophthalmol, Norwich NR4 7UZ, Norfolk, England
关键词
D O I
10.1136/bjo.2003.019331
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Aims: To investigate the effect of central corneal thickness and corneal curvature on intraocular pressure measurements using the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer, and to assess the agreement between the pulsatile ocular blood flow tonograph and the Goldmann applanation tonometer in intraocular pressure measurement. Methods: 479 subjects underwent intraocular pressure measurements with the Goldmann applanation tonometer and the pulsatile ocular blood flow tonograph. Of these, 334 patients underwent additional measurement of central corneal thickness with an ultrasonic pachymeter and corneal curvature measurement with a keratometer. Results: The intraocular pressure measurements obtained with both the Goldmann applanation tonometer and the pulsatile ocular blood flow tonograph varied with central corneal thickness and mean keratometric reading. Intraocular pressure measured using the Goldmann applanation tonometer increased by 0.027 mm Hg per mm increase in central corneal thickness. Intraocular pressure measured using the pulsatile ocular blood flow tonograph increased by 0.048 mm Hg per mm increase in central corneal thickness. For an increase of 1 mm of mean corneal curvature there was rise in intraocular pressure of 1.14 mm Hg measured by the Goldmann applanation tonometer and of 2.6 mm Hg measured by the pulsatile ocular blood flow tonograph. When compared to the Goldmann applanation tonometer, the pulsatile ocular blood flow tonograph underestimated at low intraocular pressure and overestimated at higher intraocular pressure. Conclusion: Central corneal thickness and corneal curvature affected measurements obtained with the pulsatile ocular blood flow tonograph more than they affected measurements obtained with the Goldmann applanation tonometer.
引用
收藏
页码:518 / 522
页数:5
相关论文
共 16 条
[1]   Ocular hypertension and central corneal thickness [J].
Argus, WA .
OPHTHALMOLOGY, 1995, 102 (12) :1810-1812
[2]  
Bhan A, 2002, INVEST OPHTH VIS SCI, V43, P1389
[3]   Falsely elevated intraocular pressure due to increased central corneal thickness [J].
Bron, AM ;
Creuzot-Garcher, C ;
Goudeau-Boutillon, S ;
d'Athis, P .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 1999, 237 (03) :220-224
[4]  
Copt RP, 1999, ARCH OPHTHALMOL-CHIC, V117, P14
[5]  
EHLERS N, 1975, ACTA OPHTHALMOL, V53, P34
[6]  
HANSEN FK, 1971, ACTA OPHTHALMOL, V49, P775
[7]   Central corneal thickness in normal, glaucomatous, and ocular hypertensive eyes [J].
Herndon, LW ;
Choudhri, SA ;
Cox, T ;
Damji, KF ;
Shields, MB ;
Allingham, RR .
ARCHIVES OF OPHTHALMOLOGY, 1997, 115 (09) :1137-1141
[8]   A preliminary study on the Ocular Blood Flow (OBF) of Hong Kong Chinese [J].
Lam, AKC ;
Chan, H ;
Fan, W ;
To, CH .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1999, 19 (06) :512-517
[9]   CORNEAL CURVATURE IN APPLANATION TONOMETRY [J].
MARK, HH .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1973, 76 (02) :223-224
[10]   Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic [J].
Shah, S ;
Chatterjee, A ;
Mathai, M ;
Kelly, SP ;
Kwartz, J ;
Henson, D ;
McLeod, D .
OPHTHALMOLOGY, 1999, 106 (11) :2154-2160