Self-calibrating 3D-ultrasound-based bone registration for minimally invasive orthopedic surgery

被引:95
作者
Barratt, DC
Penney, GP
Chan, CSK
Slomczykowski, M
Carter, TJ
Edwards, PJ
Hawkes, DJ
机构
[1] Kings Coll London, GKT Sch Med, Guys Hosp, Dept Imaging Sci, London SE1 9RT, England
[2] Johnson & Johnson Co, Depuy Int, Leeds LS11 0EA, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
calibration; computer-aided surgery; orthopedics; registration; ultrasound;
D O I
10.1109/TMI.2005.862736
中图分类号
TP39 [计算机的应用];
学科分类号
081203 [计算机应用技术]; 0835 [软件工程];
摘要
Intraoperative freehand three-dimensional (3-D) ultrasound (3D-US) has been proposed as a noninvasive method for registering bones to a preoperative computed tomography image or computer-generated bone model during computer-aided orthopedic surgery (CAOS). In this technique, an US probe is tracked by a 3-D position sensor and acts as a percutaneous device for localizing the bone surface. However, variations in the acoustic properties of soft tissue, such as the average speed of sound, can introduce significant errors in the bone depth estimated from US images, which limits registration accuracy. We describe a new self-calibrating approach to US-based bone registration that addresses this problem, and demonstrate its application within a standard registration scheme. Using realistic US image data acquired from 6 femurs and 3 pelves of intact human cadavers, and accurate Gold Standard registration transformations calculated using bone-implanted fiducial markers, we show that self-calibrating registration is significantly more accurate than a standard method, yielding an average root mean squared target registration error of 1.6 mm. We conclude that self-calibrating registration results in significant improvements in registration accuracy for CAOS applications over conventional approaches where calibration parameters of the 3D-US system remain fixed to values determined using a preoperative phantom-based calibration.
引用
收藏
页码:312 / 323
页数:12
相关论文
共 71 条
[1]
AHRENGART L, 1993, CLIN ORTHOP RELAT R, P153
[2]
Amin Devin V, 2003, Comput Aided Surg, V8, P1, DOI 10.3109/10929080309146097
[3]
AMIOT LP, 2004, CLIN ORTHOP RELAT R, V421, P77
[4]
A-mode ultrasound-based registration in computer-aided surgery of the skull [J].
Amstutz, C ;
Caversaccio, M ;
Kowal, J ;
Bächler, R ;
Nolte, LP ;
Häusler, R ;
Styner, M .
ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2003, 129 (12) :1310-1316
[5]
The direct estimation of sound speed using pulse-echo ultrasound [J].
Anderson, ME ;
Trahey, GE .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1998, 104 (05) :3099-3106
[6]
[Anonymous], 2001, Proc. British Machine Vision Conference
[7]
Computer-assisted distal radius osteotomy [J].
Athwal, GS ;
Ellis, RE ;
Small, CF ;
Pichora, DR .
JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2003, 28A (06) :951-958
[8]
Athwal GS, 2004, ORTHOPEDICS, V27, P195
[9]
AULT T, 1994, P 1 INT S MED ROB CO, P74
[10]
Bachler R, 2001, Comput Aided Surg, V6, P143