A highly stable biosensor for phenols prepared by immobilizing polyphenol oxidase into polyaniline-polyacrylonitrile composite matrix

被引:86
作者
Xue, HG [1 ]
Shen, ZQ [1 ]
机构
[1] Zhejiang Univ, Inst Polymer Sci, Hangzhou 310027, Peoples R China
基金
中国国家自然科学基金;
关键词
polyaniline-polyacrylonitrile composite; biosensor for phenols; operational and long-term stability; electropolymerization; polyphenol oxidase;
D O I
10.1016/S0039-9140(02)00028-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A novel biosensor for phenols was constructed by immobilizing polyphenol oxidase (PPO) into polyaniline-polyacrylonitrile composite matrix. The sensing film was prepared by electropolymerization of aniline into polyacrylonitrile (PAN)-coated platinum electrode in the presence of PPO. The scanning electron micrographs (SEM) showed that PAN had microporous structure and polyaniline and the enzyme could co-intercalated into PAN matrix. The obtained biosensor exhibited high sensitivity and excellent stability, which had no apparent loss of activity after 100 consecutive measurements and intermittent usage for 8 months with storage in a phosphate buffer at 4 C. The construction and operational conditions of the enzyme electrode were optimized. The sensitivities of the enzyme electrode for phenol, p-cresol, in-cresol and catechol were 0.96, 1.38, 1.5 and 2.03 AM(-1) cm(-2), respectively. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:289 / 295
页数:7
相关论文
共 22 条
[1]   DEPHENOLIZATION OF INDUSTRIAL WASTEWATERS CATALYZED BY POLYPHENOL OXIDASE [J].
ATLOW, SC ;
BONADONNAAPARO, L ;
KLIBANOV, AM .
BIOTECHNOLOGY AND BIOENGINEERING, 1984, 26 (06) :599-603
[2]  
BARMAN TE, 1985, ENZYME HDB, V1, P226
[3]   Co-immobilization of lactate oxidase and lactate dehydrogenase on conducting polyaniline films [J].
Chaubey, A ;
Pande, KK ;
Singh, VS ;
Malhotra, BD .
ANALYTICA CHIMICA ACTA, 2000, 407 (1-2) :97-103
[4]   Development of a PPO-poly(amphiphilic pyrrole) electrode for on site monitoring of phenol in aqueous effluents [J].
Cosnier, S ;
Fombon, JJ ;
Labbé, P ;
Limosin, D .
SENSORS AND ACTUATORS B-CHEMICAL, 1999, 59 (2-3) :134-139
[5]   Biomolecule immobilization on electrode surfaces by entrapment or attachment to electrochemically polymerized films. A review [J].
Cosnier, S .
BIOSENSORS & BIOELECTRONICS, 1999, 14 (05) :443-456
[6]   Cyro-hydrogel for the construction of a tyrosinase-based biosensor [J].
Deng, Q ;
Guo, YZ ;
Dong, SJ .
ANALYTICA CHIMICA ACTA, 1996, 319 (1-2) :71-77
[7]  
Gerard M, 1999, ELECTROANAL, V11, P450, DOI 10.1002/(SICI)1521-4109(199905)11:6<450::AID-ELAN450>3.0.CO
[8]  
2-R
[9]  
INGRID MR, 1999, ANAL CHIM ACTA, V389, P161
[10]  
KESTING RE, 1985, SYNTHETIC POLYM MEMB, P264