Clinical Differences Between Continuous Flow Ventricular Assist Devices: A Comparison Between HeartMate II and HeartWare HVAD

被引:90
作者
Lalonde, Spencer D. [1 ]
Alba, Ana C. [1 ]
Rigobon, Alanna [1 ]
Ross, Heather J. [1 ]
Delgado, Diego H. [1 ]
Billia, Filio [1 ]
McDonald, Michael [1 ]
Cusimano, Robert J. [1 ]
Yau, Terrence M. [1 ]
Rao, Vivek [1 ]
机构
[1] Toronto Gen Hosp, Univ Hlth Network, Peter Munk Cardiac Ctr, Heart Failure Transplant Program, Toronto, ON, Canada
关键词
BRIDGE; NONPULSATILE; PULSATILE;
D O I
10.1111/jocs.12158
中图分类号
R5 [内科学];
学科分类号
100201 [内科学];
摘要
Background: The HeartWare ventricular assist device (HVAD) is a new generation centrifugal flow VAD recently introduced in Canada. The objective of this study was to compare the HVAD device to the HeartMate II (HMII) axial flow device. Very few studies have compared clinical outcomes between newer generation VADs. Methods: All perioperative and follow-up data on LVAD recipients were collected prospectively in our institutional database. Between January 2006 and April 2012, 46 consecutive patients underwent implantation of either an HVAD (n=13) or a HMII (n=33) device. Pre-implant demographics, perioperative and postoperative clinical outcomes were reviewed between groups. Results: Overall, the baseline characteristics, demographics, co-morbidities and laboratory values were comparable between the two groups. The majority of the patients were Interagency Registry for Mechanical Assisted Circulatory Support 3-4 (92% in both groups) and most of the patients were bridge to transplant (75% in HMII vs. 79% in HVAD). Survival and the incidence of perioperative bleeding, renal dysfunction, liver dysfunction, and infection were similar between the groups. However, HVAD devices had a significantly higher incidence of gastrointestinal (GI) bleeding (31% vs. 0% in HMII patients, p<0.01) and stroke (44% vs. 10% in HMII patients, at one year p=0.04). Hemorrhagic strokes were more frequent in patients with HVAD (three of the five episodes vs. one of the three episodes in HMII patients, p=0.06). Conclusion: While device complications were comparable, patients with HVAD experienced a significantly higher incidence of stroke and GI bleeding and therefore refinement in patients' management may decrease incidence of these complications. doi: 10.1111/jocs.12158 (J Card Surg 2013;28:604-610)
引用
收藏
页码:604 / 610
页数:7
相关论文
共 18 条
[1]
Use of an Intrapericardial, Continuous-Flow, Centrifugal Pump in Patients Awaiting Heart Transplantation [J].
Aaronson, Keith D. ;
Slaughter, Mark S. ;
Miller, Leslie W. ;
McGee, Edwin C. ;
Cotts, William G. ;
Acker, Michael A. ;
Jessup, Mariell L. ;
Gregoric, Igor D. ;
Loyalka, Pranav ;
Frazier, O. H. ;
Jeevanandam, Valluvan ;
Anderson, Allen S. ;
Kormos, Robert L. ;
Teuteberg, Jeffrey J. ;
Levy, Wayne C. ;
Naftel, David C. ;
Bittman, Richard M. ;
Pagani, Francis D. ;
Hathaway, David R. ;
Boyce, Steven W. .
CIRCULATION, 2012, 125 (25) :3191-+
[2]
Alba Ana C, 2009, Expert Rev Cardiovasc Ther, V7, P1067, DOI 10.1586/erc.09.86
[3]
Gastrointestinal bleeding rates in recipients of nonpulsatile and pulsatile left ventricular assist devices [J].
Crow, Sheri ;
John, Ranjit ;
Boyle, Andrew ;
Shumway, Sara ;
Liao, Kenneth ;
Colvin-Adams, Monica ;
Toninato, Carol ;
Missov, Emil ;
Pritzker, Marc ;
Martin, Cindy ;
Garry, Daniel ;
Thomas, William ;
Joyce, Lyle .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2009, 137 (01) :208-215
[4]
The characteristics of adults with upper gastrointestinal bleeding admitted to Tripoli Medical Center: a retrospective case-series analysis [J].
Elghuel, Abdulbaset .
LIBYAN JOURNAL OF MEDICINE, 2011, 6 :1-5
[5]
Mechanical Circulatory Support as a Bridge to Transplant Candidacy [J].
Elhenawy, Abdelsalam M. ;
Algarni, Khaled D. ;
Rodger, Marnie ;
MacIver, Jane ;
Maganti, Manjula ;
Cusimano, Robert J. ;
Yau, Terrence M. ;
Delgado, Diego H. ;
Ross, Heather J. ;
Rao, Vivek .
JOURNAL OF CARDIAC SURGERY, 2011, 26 (05) :542-547
[6]
Clinical outcomes are similar in pulsatile and nonpulsatile left ventricular assist device recipients [J].
Feller, Erika D. ;
Sorensen, Erik N. ;
Haddad, Michel ;
Pierson, Richard N., III ;
Johnson, Frances L. ;
Brown, James M. ;
Griffith, Bartley P. .
ANNALS OF THORACIC SURGERY, 2007, 83 (03) :1082-1088
[7]
Improved survival and decreasing incidence of adverse events with the HeartMate II left ventricular assist device as bridge-to-transplant therapy [J].
John, Ranjit ;
Kamdar, Forum ;
Liao, Kenneth ;
Colvin-Adams, Monica ;
Boyle, Andrew ;
Joyce, Lyle .
ANNALS OF THORACIC SURGERY, 2008, 86 (04) :1227-1235
[8]
Time Trends and Impact of Upper and Lower Gastrointestinal Bleeding and Perforation in Clinical Practice [J].
Lanas, Angel ;
Garcia-Rodriguez, Luis A. ;
Polo-Tomas, Monica ;
Ponce, Marta ;
Alonso-Abreu, Inmaculada ;
Angeles Perez-Aisa, Maria ;
Perez-Gisbert, Javier ;
Bujanda, Luis ;
Castro, Manuel ;
Munoz, Maria ;
Rodrigo, Luis ;
Calvet, Xavier ;
Del-Pino, Dolores ;
Garcia, Santiago .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2009, 104 (07) :1633-1641
[9]
McCashland TM, 2001, AM J GASTROENTEROL, V96, P882, DOI 10.1111/j.1572-0241.2001.03638.x
[10]
Use of a continuous-flow device in patients awaiting heart transplantation [J].
Miller, Leslie W. ;
Pagani, Francis D. ;
Russell, Stuart D. ;
John, Ranjit ;
Boyle, Andrew J. ;
Aaronson, Keith D. ;
Conte, John V. ;
Naka, Yoshifumi ;
Mancini, Donna ;
Delgado, Reynolds M. ;
MacGillivray, Thomas E. ;
Farrar, David J. ;
Frazier, O. H. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (09) :885-896