Analysis of 118 second-generation metal-on-metal retrieved hip implants

被引:268
作者
Sieber, HP [1 ]
Rieker, CB [1 ]
Köttig, P [1 ]
机构
[1] Regionalspital, CH-2502 Biel, Switzerland
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1999年 / 81B卷 / 01期
关键词
D O I
10.1302/0301-620X.81B1.9047
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Osteolysis is due to particulate wear debris and is responsible for the long-term failure of total hip replacements. It has stimulated the development of alternative joint surfaces such as metal-on-metal or ceramic-on-ceramic implants, Since 1988 the second-generation metal-on-metal implant Metasul has been used in over 60 000 hips. Analysis of 118 retrieved specimens of the head or cup showed rates of wear of approximately 25 mu m for the whole articulation per year in the first year, decreasing to about 5 mu m per year after the third, Metal surfaces have a 'self-polishing' capacity. Scratches are worn out by further joint movement, Volumetric wear was decreased some 60-fold compared with that of metal-on-polyethylene implants, suggesting that second-generation metal-on-metal prostheses may considerably reduce osteolysis.
引用
收藏
页码:46 / 50
页数:5
相关论文
共 53 条
[1]   The effects of particulate cobalt, chromium and cobalt-chromium alloy on human osteoblast-like cells in vitro [J].
Allen, MJ ;
Myer, BJ ;
Millett, PJ ;
Rushton, N .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (03) :475-482
[2]  
Amstutz H. C., 1996, CLIN ORTHOP S, V329, P11
[3]  
AMSTUTZ HC, 1996, CLIN ORTHOP S, V329, P297
[4]  
BOUTIN P, 1972, REV CHIR ORTHOP, V58, P229
[5]   Elevated serum cobalt with metal-on-metal articulating surfaces [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
GottsaunerWolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :316-321
[6]  
CASE CP, 1996, CLIN ORTHOP S, V329, P269
[7]   OPTIMUM SIZE OF PROSTHETIC HEADS IN RELATION TO WEAR OF PLASTIC SOCKETS IN TOTAL REPLACEMENT OF HIP [J].
CHARNLEY, J ;
KAMANGAR, A ;
LONGFIELD, MD .
MEDICAL & BIOLOGICAL ENGINEERING, 1969, 7 (01) :31-+
[8]   Early failure of Hylamer acetabular inserts due to eccentric wear [J].
Chmell, MJ ;
Poss, R ;
Thomas, WH ;
Sledge, CB .
JOURNAL OF ARTHROPLASTY, 1996, 11 (03) :351-353
[9]  
Coleman MP, 1996, ANN CLIN LAB SCI, V26, P139
[10]  
DOORN P, 1997, 23 ANN M SOC BIOM, P192