Early failure of Hylamer acetabular inserts due to eccentric wear

被引:61
作者
Chmell, MJ [1 ]
Poss, R [1 ]
Thomas, WH [1 ]
Sledge, CB [1 ]
机构
[1] BRIGHAM & WOMENS HOSP,BOSTON,MA 02115
关键词
acetabular liners; total hip arthroplasty; failure; eccentric wear;
D O I
10.1016/S0883-5403(96)80093-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The authors performed a retrospective review of the survivorship of Hylamer (DePuy Dupont Orthopaedics, Warsaw, LN) acetabular liners following primary arthroplasty of the acetabulum. Between 1990 and 1992, 193 total hip arthroplasties were performed in which Hylamer acetabular liners were implanted. One hundred forty-three of these hips (74%) were available for clinical and radiographic follow-up evaluation after a minimum of 2 years. The mean follow-up period was 2.5 years. Five Hylamer liners have been revised for eccentric wear at a mean of 37 months postimplantation and one additional liner is scheduled to be revised at the 34-month follow-up evaluation (total, 4.2% failure rate due to wear). Survivorship of the Hylamer liner, with failure defined as revision due to eccentric wear, was calculated to be 86% (SE, 5.5%) at 4 years. The authors have discontinued the use of Hylamer acetabular liners because of this high early failure rate.
引用
收藏
页码:351 / 353
页数:3
相关论文
共 9 条
[1]   SURVIVORSHIP OF TOTAL HIP REPLACEMENTS [J].
DOBBS, HS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1980, 62 (02) :168-173
[2]   LONG-TERM RESULTS OF CHARNLEY LOW-FRICTION ARTHROPLASTY IN YOUNG-PATIENTS [J].
JOSHI, AB ;
PORTER, ML ;
TRAIL, IA ;
HUNT, LP ;
MURPHY, JCM ;
HARDINGE, K .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (04) :616-623
[3]   ULTRA-HIGH-MOLECULAR-WEIGHT POLYETHYLENE - THE MATERIAL AND ITS USE IN TOTAL JOINT IMPLANTS [J].
LI, S ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (07) :1080-1090
[4]   EFFECT OF FEMORAL-HEAD SIZE ON WEAR OF THE POLYETHYLENE ACETABULAR COMPONENT [J].
LIVERMORE, J ;
ILSTRUP, D ;
MORREY, B .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :518-528
[5]   DESIGN AND ANALYSIS OF RANDOMIZED CLINICAL-TRIALS REQUIRING PROLONGED OBSERVATION OF EACH PATIENT .2. ANALYSIS AND EXAMPLES [J].
PETO, R ;
PIKE, MC ;
ARMITAGE, P ;
BRESLOW, NE ;
COX, DR ;
HOWARD, SV ;
MANTEL, N ;
MCPHERSON, K ;
PETO, J ;
SMITH, PG .
BRITISH JOURNAL OF CANCER, 1977, 35 (01) :1-39
[6]   THE RELATIONSHIP BETWEEN THE DESIGN, POSITION, AND ARTICULAR WEAR OF ACETABULAR COMPONENTS INSERTED WITHOUT CEMENT AND THE DEVELOPMENT OF PELVIC OSTEOLYSIS [J].
SCHMALZRIED, TP ;
GUTTMANN, D ;
GRECULA, M ;
AMSTUTZ, HC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1994, 76A (05) :677-688
[7]   PERIPROSTHETIC BONE LOSS IN TOTAL HIP-ARTHROPLASTY - POLYETHYLENE WEAR DEBRIS AND THE CONCEPT OF THE EFFECTIVE JOINT SPACE [J].
SCHMALZRIED, TP ;
JASTY, M ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (06) :849-863
[8]   PATTERNS OF OSTEOLYSIS AROUND TOTAL HIP COMPONENTS INSERTED WITH AND WITHOUT CEMENT [J].
ZICAT, B ;
ENGH, CA ;
GOKCEN, E .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77A (03) :432-439
[9]  
1994, TEST DATA RESULTS HY