Comparison of the performance characteristics of three generations of membrane oxygenators: Univox®, Univox® Gold™ and SpiralGold™

被引:5
作者
Beckley, Philip D. [1 ]
Morris, Susan M. [1 ]
Smith, James J. [1 ]
McNamara, Jerri L. [1 ]
Novak, Julie A. [1 ]
机构
[1] Ohio State Univ, Circulat Technol Div, Columbus, OH 43210 USA
来源
PERFUSION-UK | 1996年 / 11卷 / 01期
关键词
D O I
10.1177/026765919601100109
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
With continuous enhancement in oxygenator design, the question is raised as to how these changes actually impact the performance of the oxygenator. The recent addition of two new oxygenators by the Bentley Division of Baxter Healthcare Corporation provided us with a unique opportunity to compare the performance of each device and isolate the impact of each design change on performance. While the basic design and flow patterns have remained the same, application of the Duraflo (R) II treatment has produced the Univox (R) Gold (TM) and a change in the fibre-winding technique has produced the SpiralGold (TM). This study compared the effects of heparin coating (Univox (R) to Univox (R) Gold (TM)) and fibre-winding (Univox (R) Gold (TM) to SpiralGold (TM);) on gas and heat transfer and resistance to blood flow (pressure drop). Six oxygenators of each model were evaluated utilizing an in vitro single pass circuit, which first conditioned bovine blood to the Association for the Advancement of Medical Instrumentation (AAMI) venous standards. Blood flows of 4.0, 5.0, 6.0 and 7.0 I/min, FiO(2) values of 1.0, 0.8 and 0.6, and gas-to-blood flow ratios of 0.5, 1.0 and 1.5 were chosen as test variables. Data generated included oxygen transfer, carbon dioxide transfer, arterial pO(2) resistance to blood flow, and coefficient of heat exchange. The results indicate that the Duraflo II treatment does not have a significant effect on gas and heat transfer or resistance to blood flow. The fibre-winding technique employed with the new SpiralGold (TM), however, has improved significantly gas exchange and arterial PO2 when compared with the previous Univox (R) models. Resistance to blood flow and coefficient of heat exchange were not affected significantly by the winding technique.
引用
收藏
页码:61 / 70
页数:10
相关论文
共 17 条
[1]  
Association for the Advancement of Medical Instrumentation, 71992 AAMIISO
[2]  
Bethune D. W., 1994, Perfusion, V9, P207, DOI 10.1177/026765919400900308
[3]  
BOONSTRA P W, 1992, Perfusion (London), V7, P109, DOI 10.1177/026765919200700205
[4]  
Cherkas DD, 1992, ANESTEZIOL REANIMAT, V4, P56
[5]   ASSESSMENT OF OXYGEN-TRANSFER IN MEMBRANE OXYGENATORS DURING CLINICAL CARDIOPULMONARY BYPASS [J].
CLAYTON, RH ;
PEARSON, DT ;
MURRAY, A .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1992, 13 (02) :167-177
[6]  
Courtney James M., 1994, Perfusion, V9, P3, DOI 10.1177/026765919400900102
[7]  
FRIED D W, 1992, Perfusion (London), V7, P141, DOI 10.1177/026765919200700210
[8]  
Gaylor John D. S., 1994, Perfusion, V9, P173, DOI 10.1177/026765919400900304
[9]  
Griffith K E, 1994, J Extra Corpor Technol, V26, P114
[10]  
Gu YJ, 1991, PERFUSION, V6, P221