Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking

被引:76
作者
Farr, J. B. [1 ,2 ]
Mascia, A. E. [2 ]
Hsi, W. -C. [2 ]
Allgower, C. E. [2 ]
Jesseph, F. [2 ]
Schreuder, A. N. [2 ]
Wolanski, M. [2 ]
Nichiporov, D. F. [3 ]
Anferov, V. [3 ]
机构
[1] Indiana Univ, Dept Phys, Bloomington, IN 47405 USA
[2] Midwest Proton Radiotherapy Inst, Bloomington, IN 47408 USA
[3] Indiana Univ, Cyclotron Facil, Bloomington, IN 47408 USA
基金
美国国家卫生研究院;
关键词
proton; commissioning; uniform; scanning;
D O I
10.1118/1.2982248
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
A proton beam delivery system on a gantry with continuous uniform scanning and dose layer stacking at the Midwest Proton Radiotherapy Institute has been commissioned and accepted for clinical use. This paper was motivated by a lack of guidance on the testing and characterization for clinical uniform scanning systems. As such, it describes how these tasks were performed with a uniform scanning beam delivery system. This paper reports the methods used and important dosimetric characteristics of radiation fields produced by the system. The commissioning data include the transverse and longitudinal dose distributions, penumbra, and absolute dose values. Using a 208 MeV cyclotron's proton beam, the system provides field sizes up to 20 and 30 cm in diameter for proton ranges in water up to 27 and 20 cm, respectively. The dose layer stacking method allows for the flexible construction of spread-out Bragg peaks with uniform modulation of up to 15 cm in water, at typical dose rates of 1-3 Gy/min. For measuring relative dose distributions, multielement ion chamber arrays, small-volume ion chambers, and radiographic films were employed. Measurements during the clinical commissioning of the system have shown that the lateral and longitudinal dose uniformity of 2.5% or better can be achieved for all clinically important field sizes and ranges. The measured transverse penumbra widths offer a slight improvement in comparison to those achieved with a double scattering beam spreading technique at the facility. Absolute dose measurements were done using calibrated ion chambers, thermoluminescent and alanine detectors. Dose intercomparisons conducted using various types of detectors traceable to a national standards laboratory indicate that the measured dosimetry data agree with each other within 5%. (C) 2008 American Association of Physicists in Medicine. [DOI: 10.1118/1.2982248]
引用
收藏
页码:4945 / 4954
页数:10
相关论文
共 38 条
[1]
Combined X-Y scanning magnet for conformal proton radiation therapy [J].
Anferov, V .
MEDICAL PHYSICS, 2005, 32 (03) :815-818
[2]
Anferov V, 2001, AIP CONF PROC, V600, P27, DOI 10.1063/1.1435188
[3]
Physical specifications of clinical proton beams from a synchrotron [J].
Arduini, G ;
Cambria, R ;
Canzi, C ;
Gerardi, F ;
Gottschalk, B ;
Leone, R ;
Sangaletti, L ;
Silari, M .
MEDICAL PHYSICS, 1996, 23 (06) :939-951
[4]
CHU W, 1993, 33749 LBL U CAL DAV
[5]
CORAY D, 2000, INT S STAND COD PRAC, P213
[6]
DeLuca P., 2007, J ICRU, V7, P29
[7]
Comparison of intensity modulated x-ray therapy and intensity modulated proton therapy for selective subvolume boosting: a phantom study [J].
Flynn, R. T. ;
Barbee, D. L. ;
Mackie, T. R. ;
Jeraj, R. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (20) :6073-6091
[8]
FRIESEL DL, 2006, EUR PART ACC C ED UK, P2349
[9]
Range modulation in proton therapy - an optimization technique for clinical and experimental applications [J].
Gardey, KU ;
Oelfke, U ;
Lam, GKY .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (06) :N81-N88
[10]
HAISEN L, 2008, MED PHYS, V35, P1103, DOI DOI 10.1118/1.2836954