Dimensions of central venous structures in humans measured in vivo using magnetic resonance imaging:: Implications for central-vein catheter dimensions

被引:16
作者
Twardowski, ZJ
Seger, RM
机构
[1] Univ Missouri, Dept Med, Columbia, MO USA
[2] Univ Missouri, Dept Radiol, Columbia, MO USA
关键词
central veins; central vein catheter; MRI; hemodialysis;
D O I
10.1177/039139880202500205
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The tip of a central vein catheter for hemodialysis should be located in the upper right atrium for the best performance. Hemodialysis catheters do have internal diameters unadjusted to the catheter length; however, the longer the catheter the slower the flow at the same pressure difference. On the other hand, the catheter diameter cannot be so large as to fill the vein too tightly as it predisposes to damage of the vein wall, thrombosis and stenosis. Therefore, the catheter length and diameter should be appropriate for the patient. For this purpose, the exact dimensions of the venous system in vivo should be known. In this study we correlated the anthropometric measurements and the dimensions of the large upper body veins in 31 adult volunteers. After deep inspiration, magnetic resonance imaging of the chest was performed in three planes; the positions of specific points in the three-dimensional coordinate system were measured, and the distance to adjacent points was calculated according to the analytic geometry formula. The total length from the catheter entry point to the right atrium was the sum of distances between the adjacent points. The lengths of the veins were correlated with the body anthropometric measurements (height, weight, body surface area, bi-acromion span, and height plus bi-acromion span). The best overall correlations of the lengths and diameters of the large upper body veins are with the body surface area. A table is included to guide the selection of the total catheter length and diameter in relation to the body surface area and insertion site.
引用
收藏
页码:107 / 123
页数:17
相关论文
共 11 条
[1]  
Davenport Andrew, 2000, Hemodial Int, V4, P78, DOI 10.1111/hdi.2000.4.1.78
[2]  
DIXIT A, 1999, J AM SOC NEPHROL, V10, pA1042
[3]   Right atrial thrombi are related to indwelling central venous catheter position: Insights into time course and possible mechanism of formation [J].
Gilon, D ;
Schechter, D ;
Rein, AJJT ;
Gimmon, Z ;
Or, R ;
Rozenman, Y ;
Slavin, S ;
Gotsman, MS ;
Nagler, A .
AMERICAN HEART JOURNAL, 1998, 135 (03) :457-462
[4]  
Jean G, 1997, NEPHROL DIAL TRANSPL, V12, P1689
[5]   Right atrial thrombi in children with cancer and indwelling catheters [J].
Korones, DN ;
Buzzard, CJ ;
Asselin, BL ;
Harris, JP .
JOURNAL OF PEDIATRICS, 1996, 128 (06) :841-846
[6]  
MCLOUD TC, 1998, THORACIC RADIOLOGY R, P1
[7]   The hemodialysis catheter conundrum: Hate living with them, but can't live without them [J].
Schwab, SJ ;
Beathard, G .
KIDNEY INTERNATIONAL, 1999, 56 (01) :1-17
[8]   All currently used measurements of recirculation in blood access by chemical methods are flawed due to intradialytic disequilibrium or recirculation at low flow [J].
Twardowski, ZJ ;
Van Stone, JC ;
Haynie, JD .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1998, 32 (06) :1046-1058
[9]  
TWARDOWSKI ZJ, 2001, J VASCULAR ACCESS, V2, P17
[10]   TOTAL-BODY WATER VOLUMES FOR ADULT MALES AND FEMALES ESTIMATED FROM SIMPLE ANTHROPOMETRIC MEASUREMENTS [J].
WATSON, PE ;
WATSON, ID ;
BATT, RD .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1980, 33 (01) :27-39