A total of 102 revision total knee arthroplasties (TKAs) were reviewed to determine the prodromal symptoms and radiographic findings associated with failure. Presenting symptoms included pain (84%); swelling (76%); progressive varus or valgus deformity (19%); instability (17%); stiffness (17%); clicking or grinding (7%); catching (4%); and patellar pain, subluxation, or clicking (4%). Radiographs were diagnostic in 91% of cases, demonstrating complete radiolucencies (80%), polyethylene wear (43%), component breakage (5%), metallic debris (3%), patellar subluxation or dislocation (4%), and osteolysis (4%). Gross intraoperative findings included polyethylene wear (72%), osteolysis (22%), metal-wear synovitis (9%), component breakage (6%), patellar wear and dissociation (4%), and occult sepsis (5%). The average duration of symptoms was 13 months; the interval between orthopedic evaluations averaged 23 months. Based on this information, we recommend that an annual questionnaire and weight-bearing radiographs be used to ensure adequate surveillance of TKA patients and avoid complications associated with delay in diagnosis of polyethylene wear or implant loosening.
引用
收藏
页码:488 / 492
页数:5
相关论文
共 15 条
[1]   POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[2]   Current concepts review managed care: Form, function, and evolution [J].
Gottlieb, S ;
Einhorn, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (01) :125-136
[3]   Quality assurance in capitated physician groups - Where is the emphasis? [J].
Kerr, EA ;
Mittman, BS ;
Hays, RD ;
Leake, B ;
Brook, RH .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (15) :1236-1239
[4]   Tibial polyethylene failure after primary porous-coated anatomic total knee arthroplasty - Aids to diagnosis and revision [J].
Knight, JL ;
Gorai, PA ;
Atwater, RD ;
Grothaus, L .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :748-757
[5]  
Landy MM, 1988, J ARTHROPLASTY S, VS3, P73
[6]  
MCGINTY JB, 1996, ORTHOPEDICS TODA JUN, P4
[7]  
MICHAELS DM, 1995, MED GROUP MANAGE MAY, P41
[8]  
RANAWAT CS, 1993, CLIN ORTHOP RELAT R, P94
[9]  
RITTER MA, 1994, CLIN ORTHOP RELAT R, P136
[10]  
RORABECK CH, 1995, ORTHOPEDICS, V18, P915