Pump Replacement for Left Ventricular Assist Device Failure Can Be Done Safely and Is Associated With Low Mortality

被引:97
作者
Moazami, Nader [1 ]
Milano, Carmelo A.
John, Ranjit
Sun, Benjamin
Adamson, Robert M.
Pagani, Francis D.
Smedira, Nicholas
Slaughter, Mark S.
Farrar, David J.
Frazier, O. Howard
机构
[1] Cleveland Clin, Kaufman Ctr Heart Failure Cardiac Transplantat &, Dept Thorac & Cardiovasc Surg, Cleveland, OH 44195 USA
关键词
HEART-FAILURE; AORTIC-INSUFFICIENCY; CIRCULATORY SUPPORT; COMMISSURAL FUSION; EXCHANGE; XVE; TRANSPLANTATION; OUTCOMES;
D O I
10.1016/j.athoracsur.2012.09.011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Although continuous-flow left ventricular assist devices (LVAD) are durable and reliable, device replacement will be inevitable in some patients. We evaluated the incidence and outcomes of pump replacement procedures with the HeartMate II (Thoratec Corporation, Pleasanton, CA) LVAD. Methods. Data were obtained from 1,128 patients implanted from March 2005 to January 2010 with the HeartMate II during the clinical trials for bridge to transplant and destination therapy. The operative mortality associated with the replacement procedure was determined. Results. The mean duration of HeartMate II support was 568 +/- 535 days (cumulative duration: 1,755 patient-years, longest: 6.5 years). A total of 72 (6.4%) patients underwent 79 LVAD replacements (0.045 events/patient-year) of which 2 were in the initial operation and 77 in separate procedures. Reasons for replacement were percutaneous lead damage (36 events, 3.0%), device thrombosis (25 events, 2.1%), infection (7 events, 0.6%), and miscellaneous other (11 events, 0.9%). The median time to pump replacement was 428 days (range 0 to 1,474). Of the 77 replacement procedures, there were 5 (6.5%) operative deaths within 30 days. The causes of death were device thrombosis, right heart failure, multisystem organ failure, and bleeding. One year after exchange (median 2.1 years after initial implant), 30% had died, 5% were transplanted, and 65% were ongoing and alive. Conclusions. HeartMate II device failure requiring pump replacement is infrequent, but when required can be done safely. These data continue to provide encouraging evidence supporting HeartMate II use for long-term circulatory support. (Ann Thorac Surg 2013;95:500-5) (c) 2013 by The Society of Thoracic Surgeons
引用
收藏
页码:500 / 505
页数:6
相关论文
共 24 条
[1]  
Banchs Javier E, 2006, J Am Soc Echocardiogr, V19, DOI 10.1016/j.echo.2006.05.025
[2]   Low Thromboembolism and Pump Thrombosis With the HeartMate II Left Ventricular Assist Device: Analysis of Outpatient Anti-coagulation [J].
Boyle, Andrew J. ;
Russell, Stuart D. ;
Teuteberg, Jeffrey J. ;
Slaughter, Mark S. ;
Moazami, Nader ;
Pagani, Francis D. ;
Frazier, O. Howard ;
Heatley, Gerald ;
Farrar, David J. ;
John, Ranjit .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2009, 28 (09) :881-887
[3]   Electrophysiologic characteristics and catheter ablation of ventricular tachyarrhythmias among patients with heart failure on ventricular assist device support [J].
Cantillon, Daniel J. ;
Bianco, Christopher ;
Wazni, Oussama M. ;
Kanj, Mohamed ;
Smedira, Nicholas G. ;
Wilkoff, Bruce L. ;
Starling, Randall C. ;
Saliba, Walid I. .
HEART RHYTHM, 2012, 9 (06) :859-864
[4]   Acquired commissural fusion of aortic valves in patients with left ventricular assist devices [J].
Connelly, JH ;
Abrams, J ;
Klima, T ;
Vaughn, WK ;
Frazier, OH .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2003, 22 (12) :1291-1295
[5]   The Development of Aortic Insufficiency in Left Ventricular Assist Device-Supported Patients [J].
Cowger, Jennifer ;
Pagani, Francis D. ;
Haft, Jonathan W. ;
Romano, Matthew A. ;
Aaronson, Keith D. ;
Kolias, Theodore J. .
CIRCULATION-HEART FAILURE, 2010, 3 (06) :668-674
[6]   Clinical experience with sternotomy versus subcostal approach for exchange of the HeartMate XVE to the HeartMate II ventricular assist device [J].
Gregoric, Igor D. ;
Bruckner, Brian A. ;
Jacob, Leon ;
Kar, Biswajit ;
Cohn, William E. ;
La Francesca, Saverio ;
Frazier, O. H. .
ANNALS OF THORACIC SURGERY, 2008, 85 (05) :1646-1650
[7]   Exchange techniques for implantable ventricular assist devices [J].
Gregoric, Igor D. .
ASAIO JOURNAL, 2008, 54 (01) :14-19
[8]   Less Frequent Opening of the Aortic Valve and a Continuous Flow Pump Are Risk Factors for Postoperative Onset of Aortic Insufficiency in Patients With a Left Ventricular Assist Device [J].
Hatano, Masaru ;
Kinugawa, Koichiro ;
Shiga, Taro ;
Kato, Naoko ;
Endo, Miyoko ;
Hisagi, Motoyuki ;
Nishimura, Takashi ;
Yao, Atsushi ;
Hirata, Yasunobu ;
Kyo, Shunei ;
Ono, Minoru ;
Nagai, Ryozo .
CIRCULATION JOURNAL, 2011, 75 (05) :1147-1155
[9]   Urgent Exchange of a HeartMate II Left Ventricular Assist Device After Percutaneous Lead Fracture [J].
Jafar, Mehdi ;
Gregoric, Igor D. ;
Radovancevic, Rajko ;
Cohn, William E. ;
McGuire, Nichole ;
Frazier, O. H. .
ASAIO JOURNAL, 2009, 55 (05) :523-524
[10]   Continuous Flow Left Ventricular Assist Device Outcomes in Commercial Use Compared With the Prior Clinical Trial [J].
John, Ranjit ;
Naka, Yoshifumi ;
Smedira, Nicholas G. ;
Starling, Randall ;
Jorde, Ulrich ;
Eckman, Peter ;
Farrar, David J. ;
Pagani, Francis D. .
ANNALS OF THORACIC SURGERY, 2011, 92 (04) :1406-1412