Precision measurement of the magnetic field In terms of the free-proton NMR frequency

被引:56
作者
Fei, X [1 ]
Hughes, VW [1 ]
Prigl, R [1 ]
机构
[1] YALE UNIV,DEPT PHYS,NEW HAVEN,CT 06511
关键词
D O I
10.1016/S0168-9002(97)84161-7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have developed a NMR standard probe using a spherical pure water sample of 1 cm diameter to determine the absolute magnetic field B in terms of the free-proton NMR frequency f(p) with an accuracy of 3.4 x 10(-8). Our standard probe can be used conveniently to calibrate other probes in the field range from 1.45 to 1.7T and can readily be employed over a much wider field range. The probe design and the tests carried out to verify its precision and accuracy are presented.
引用
收藏
页码:349 / 356
页数:8
相关论文
共 29 条
[21]   MAGNETIC-MOMENT OF PROTON IN H2O IN BOHR MAGNETONS [J].
PHILLIPS, WD ;
COOKE, WE ;
KLEPPNER, D .
METROLOGIA, 1977, 13 (04) :179-195
[22]   A high precision magnetometer based on pulsed NMR [J].
Prigl, R ;
Haeberlen, U ;
Jungmann, K ;
Putlitz, GZ ;
vonWalter, P .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 374 (01) :118-126
[23]  
PRIGL R, 1993, 161 BROOKH NAT LAB
[24]  
ROBERTS L, 1992, FUTURE MUON PHYSICS, P101
[25]  
SCHETGEN R, 1993, ARRL HDB RADIO AMATE
[26]   DIFFUSION AND NUCLEAR SPIN RELAXATION IN WATER [J].
SIMPSON, JH ;
CARR, HY .
PHYSICAL REVIEW, 1958, 111 (05) :1201-1202
[27]  
WEAST RC, 1986, CRC HDB CHEM PHYSICS
[28]   MAGNETIC MOMENT AND HFS ANOMALY FOR HE3 [J].
WILLIAMS, WL ;
HUGHES, VW .
PHYSICAL REVIEW, 1969, 185 (04) :1251-&
[29]  
[No title captured]