A prototype of ultrasonic micro-degassing device for portable dialysis system

被引:38
作者
Yang, Z
Matsumoto, S
Maeda, R
机构
[1] AIST, Inst Engn Mech Syst, Tsukuba, Ibaraki 3058564, Japan
[2] NEDO, Tokyo 1706028, Japan
关键词
degassing; dialysis; ultrasound; micromachine;
D O I
10.1016/S0924-4247(01)00741-5
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The design, fabrication and evaluation of a micro-degasser are described. The intended use of the device is in portable dialysis systems, Degassing processes were based on ultrasound induced cavitation. The pattern of the degassing chamber was formed in a glass wafer and that of the gas-venting channels were formed in a silicon wafer. The entire flow path network was encapsulated by the anodic bonding of the Si to the glass wafer. A diaphragm (6 mm x 6 mm x 0.1 mm) was etched on the Si side for oscillation. The ultrasonic vibration originated from a bulk piezoelectric PZT ceramic (5 mm x 4 mm x 0.15 mm) which was excited by a 49 kHz square wave at 100 V (peak-to-peak). The gas-venting channels (2 mum wide and 2.7 mum deep) were hydrophobically passivated using a functional silicone. Cavitation occurred in a degassing chamber (6 mm x 6 mm x 0.02 mm) when the Si oscillating diaphragm was driven by the PZT. Water was used to demonstrate the degassing process. The entire process was recorded using a microscope equipped with a video camera. The gas bubbles were vented effectively and no gas bubble flowing out of the degassing chamber with water was observed. Thirty-eight percent of the dissolved oxygen was removed. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:274 / 280
页数:7
相关论文
共 10 条
[1]   MICROMACHINING OF CAPILLARY ELECTROPHORESIS INJECTORS AND SEPARATORS ON GLASS CHIPS AND EVALUATION OF FLOW AT CAPILLARY INTERSECTIONS [J].
FAN, ZH ;
HARRISON, DJ .
ANALYTICAL CHEMISTRY, 1994, 66 (01) :177-184
[2]  
FROST TH, 1978, TECHNICAL ASPECTS RE, P42
[3]  
HITCHMAN ML, 1978, MEASUREMENT DISSOLVE, P59
[4]  
Hosokawa K., 1998, MICRO TOTAL ANAL SYS, V98, P307
[5]   A self-priming and bubble-tolerant piezoelectric silicon micropump for liquids and gases [J].
Linnemann, R ;
Woias, P ;
Senfft, CD ;
Ditterich, JA .
MICRO ELECTRO MECHANICAL SYSTEMS - IEEE ELEVENTH ANNUAL INTERNATIONAL WORKSHOP PROCEEDINGS, 1998, :532-537
[6]  
OOTA T, 1993, Patent No. 06293
[7]  
OTA K, 1979, DIALYSIS TREATMENT R, P67
[8]  
*TRC R D LIB, 1996, REP REM SUPPL DISS G, P7
[9]   Fabrication of oxygen electrode arrays and their incorporation into sensors for measuring biochemical oxygen demand [J].
Yang, Z ;
Sasaki, S ;
Karube, I ;
Suzuki, H .
ANALYTICA CHIMICA ACTA, 1997, 357 (1-2) :41-49
[10]  
Yang Z, 2000, MEMS 2000, P80