Radiation dose to the fetus from body MDCT during early gestation.

被引:172
作者
Hurwitz, LM
Yoshizumi, T
Reiman, RE
Goodman, PC
Paulson, EK
Frush, DP
Toncheva, G
Nguyen, G
Barnes, L
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
[2] Duke Univ, Radiat Safety Div, Durham, NC 27710 USA
关键词
CT; fetal imaging; radiation dose; women's imaging;
D O I
10.2214/AJR.04.1915
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The purpose of our study was to determine radiation dose to the fetus at early gestation when contemporary MDCT scanners are used for common clinical indications. MATERIALS AND METHODS. Anthropomorphic phantoms were constructed to reflect a pregnant woman. Thermoluminescence dosimeters (TLDs) and metal oxide semiconductor field effect transistor (MOSFET) detectors were placed in appropriate locations to determine real-time radiation exposure to the fetus at 0 and 3 months' Gestation. Imaging was performed on a 16-MDCT scanner using current institutional CT protocols: renal stone (140 kVp, 160 mA, rotation time of 0.5 see, 16 x 0.625 mm), appendix (140 kVp, 340 mA, rotation time of 0.5 see, 16 x 0.625 mm), and pulmonary embolus (140 kVp 380 mA. rotation time of 0.8 sec, 16 x 1.25 mm). RESULTS. The radiation dose to the fetus at 0 and 3 months, respectively.. was as follows: renal stone protocol, 0.8-1.2 and 0.4-0.7 cGy; appendix protocol, 1.52-1.68 and 2-4 cGy; and pulmonary embolus protocol, 0.024-0.047 and 0.061-0.066 cGy. CONCLUSION. Radiation doses to the fetus from institutional MDCT protocols that may be used during pregnancy (for pulmonary embolus, appendicitis, and renal colic) are below the level thought to induce neurologic detriment to the fetus. Imaging the mother for appendicitis theoretically may double the fetal risk for developing a childhood cancer. Radiation doses to the fetus from pulmonary embolus chest CT angiography are of the same magnitude as ventilation-perfusion (V/Q) scanning.
引用
收藏
页码:871 / 876
页数:6
相关论文
共 36 条
[1]  
AGNER LK, 1997, EXPOSURE PREGNANT PA, P93
[2]  
[Anonymous], 100 NCRP
[3]  
BETTMAN MA, 1909, ACR APPROPRIATENESS, P1
[4]   Pulmonary embolus pregnant patients: Survey of ventilation-perfusion imaging policies and practices [J].
Boiselle, PM ;
Reddy, SS ;
Villas, PA ;
Liu, A ;
Seibyl, JP .
RADIOLOGY, 1998, 207 (01) :201-206
[5]   The use of helical computed tomography in pregnancy for the diagnosis of acute appendicitis [J].
Castro, MA ;
Shipp, TD ;
Castro, EE ;
Ouzounian, J ;
Rao, P .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2001, 184 (05) :954-957
[6]   Trends in the use of unenhanced helical CT for acute urinary colic [J].
Chen, MYM ;
Zagoria, RJ ;
Saunders, HS ;
Dyer, RB .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1999, 173 (06) :1447-1450
[7]   MRI for clinically suspected appendicitis during pregnancy [J].
Cobben, LP ;
Groot, I ;
Haans, L ;
Blickman, JG ;
Puylaert, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (03) :671-675
[8]   Estimation of fetal radiation dose from computed tomography scanning in late pregnancy - Depth-dose data from routine examinations [J].
Damilakis, J ;
Perisinakis, K ;
Voloudaki, A ;
Gourtsoyiannis, N .
INVESTIGATIVE RADIOLOGY, 2000, 35 (09) :527-533
[9]   ESTIMATED FETAL RADIATION-DOSE FROM MULTISLICE CT STUDIES [J].
FELMLEE, JP ;
GRAY, JE ;
LEETZOW, ML ;
PRICE, JC .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1990, 154 (01) :185-190
[10]  
GILMAN E A, 1988, Journal of Radiological Protection, V8, P3, DOI 10.1088/0952-4746/8/1/301