Welder identity, weld date, and the risk of outlet strut fracture in Bjork-Shiley convexo-concave valves: The Dutch cohort study

被引:9
作者
Kallewaard, M
Algra, A
vanderGraaf, Y
机构
[1] Department of Clinical Epidemiology, School of Medicine, Utrecht University
[2] Clinical Epidemiology Unit, School of Medicine, Utrecht University, 3508 TA Utrecht
关键词
heart valve prosthesis; welder identity; strut fracture; Bjork-Shiley CC valve;
D O I
10.1136/hrt.76.6.510
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective-To establish whether there is an association between subsequent fracture of the outlet strut and welder identity and the weld date of Bjork-Shiley convexo-concave (BScc) valves. Design-Cohort study. Patients-All Dutch BScc valve recipients (n = 2266). Mean outcome measures-Documented outlet strut fracture during follow up. Results-Weld dates were known for 97.0% of all BScc valves (n = 2534) implanted in Dutch patients (n = 2266) and welder identity was known for 52.2%. During a mean follow up of 9.4 years, 46 fractures were documented. For 60 degrees valves welded from 1981 to 1984 the fracture rate (0.22 per 100 person-years (95% CI 0.13 to 0.34)) was higher than that for valves welded before 1981 (0.04 (95% CI 0.01-0.10)). When all fracture related risk factors were taken into account, fracture rates per welder did not show any statistically significant differences. Conclusions-Welder identity does not contribute to the risk of strut fracture. Other factors in the production of BScc valves may explain the increased risk for valves welded from 1981 to 1984.
引用
收藏
页码:510 / 512
页数:3
相关论文
共 9 条
[1]   SINGLE-LEG STRUT FRACTURES IN EXPLANTED BJORK-SHILEY VALVES [J].
DEMOL, BA ;
KALLEWAARD, M ;
MCLELLAN, RB ;
VANHERWERDEN, LA ;
DEFAUW, JJ ;
VANDERGRAAF, Y .
LANCET, 1994, 343 (8888) :9-12
[2]  
HEDGER P, 1995, IMPORTANT UPDATE INF
[3]   OUTLET STRUT FRACTURE OF THE BJORK-SHILEY 60-DEGREES CONVEXO-CONCAVE VALVE - CURRENT INFORMATION AND RECOMMENDATIONS FOR PATIENT-CARE [J].
HIRATZKA, LF ;
KOUCHOUKOS, NT ;
GRUNKEMEIER, GL ;
MILLER, DC ;
SCULLY, HE ;
WECHSLER, AS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (05) :1130-1137
[4]  
LINDBLOM D, 1989, J THORAC CARDIOV SUR, V97, P95
[5]   THE BJORK-SHILEY 70-DEGREES CONVEXO-CONCAVE PROSTHESIS STRUT FRACTURE PROBLEM - (PRESENT STATE OF INFORMATION) [J].
OSTERMEYER, J ;
HORSTKOTTE, D ;
BENNETT, J ;
HUYSMANS, H ;
LINDBLOM, D ;
OLIN, C ;
ORINIUS, E ;
SEMB, G .
THORACIC AND CARDIOVASCULAR SURGEON, 1987, 35 (02) :71-77
[6]  
*POISS VERT, 1960, DOC GEIG WISS TAB, P107
[7]   LETHAL BLOCKAGE OF A BJORK-SHILEY ARTIFICIAL-HEART VALVE CAUSED BY STRUT FRACTURE - THE METALLURGICAL ASPECT [J].
ROCKELEIN, G ;
BREME, J ;
VONDEREMDE, J .
THORACIC AND CARDIOVASCULAR SURGEON, 1989, 37 (01) :47-51
[8]   METALLURGICAL ANALYSIS OF FAILED BJORK-SHILEY CARDIAC-VALVE PROSTHESES [J].
SACKS, SH ;
HARRISON, M ;
BISCHLER, PJE ;
MARTIN, JW ;
WATKINS, J ;
GUNNING, ALF .
THORAX, 1986, 41 (02) :142-147
[9]   RISK OF STRUT FRACTURE OF BJORK-SHILEY VALVES [J].
VANDERGRAAF, Y ;
DEWAARD, F ;
VANHERWERDEN, LA ;
DEFAUW, J .
LANCET, 1992, 339 (8788) :257-261