SINGLE-LEG STRUT FRACTURES IN EXPLANTED BJORK-SHILEY VALVES

被引:34
作者
DEMOL, BA
KALLEWAARD, M
MCLELLAN, RB
VANHERWERDEN, LA
DEFAUW, JJ
VANDERGRAAF, Y
机构
[1] DELFT UNIV TECHNOL,SAFETY SCI GRP,DELFT,NETHERLANDS
[2] UNIV UTRECHT,DEPT EPIDEMIOL,UTRECHT,NETHERLANDS
[3] CARDIOPULM SURG CTR,AMSTERDAM,NETHERLANDS
[4] ST ANTONIUS HOSP,DEPT CARDIOPULM SURG,NIEUWEGEIN,NETHERLANDS
[5] RICE UNIV,DEPT MECH ENGN & MAT SCI,HOUSTON,TX 77251
[6] ERASMUS UNIV ROTTERDAM,HOSP DIJKZIGT,DEPT CARDIOPULM SURG,ROTTERDAM,NETHERLANDS
关键词
D O I
10.1016/S0140-6736(94)90874-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A retrospective follow-up study in the Netherlands on the risk of fracture of Bjork-Shiley convexoconcave valves concluded that prophylactic replacement is advisable for certain groups of patients. We have examined valves explanted from twenty-two patients because they met epidemiological criteria for reoperation, with or without the presence of moderate cardiac impairment, or because there were other cardiac complaints not related to the valve. No information was available before explantation to suggest a valve defect. All patients survived their operations. Of the twenty-four valves, seven (29% [95% CI 13-52]) had fracture of one of the legs of the outlet strut (single-leg strut fracture [SLF]). Two other valves had features that suggested fatigue defects. As in the previous study, 70-degrees valves had the highest risk of SLF (five of sixteen, two aortic and three mitral). However, two of eight 60-degrees valves (both aortic) also had SLF. Current hazard calculations and explantation recommendations may need to be revised. Since several of the defective valves were welded by the same person, knowledge of manufacturing details may help in estimation of fracture risk for an individual patient.
引用
收藏
页码:9 / 12
页数:4
相关论文
共 14 条
[1]   SHOULD PATIENTS WITH BJORK-SHILEY VALVES UNDERGO PROPHYLACTIC REPLACEMENT [J].
BIRKMEYER, JD ;
MARRIN, CAS ;
OCONNOR, GT .
LANCET, 1992, 340 (8818) :520-523
[2]   METALLURGIC AND DESIGN DEVELOPMENT IN RESPONSE TO MECHANICAL DYSFUNCTION OF BJORK-SHILEY HEART-VALVES [J].
BJORK, VO .
SCANDINAVIAN JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1985, 19 (01) :1-12
[3]  
Blackstone E H, 1992, J Heart Valve Dis, V1, P3
[4]   STRUT FRACTURE WITH BJORK-SHILEY 70-PERCENT CONVEXO-CONCAVE VALVE - AN INTERNATIONAL MULTIINSTITUTIONAL FOLLOW-UP-STUDY [J].
ERICSSON, A ;
LINDBLOM, D ;
SEMB, G ;
HUYSMANS, HA ;
THULIN, LI ;
SCULLY, HE ;
BENNETT, JG ;
OSTERMEYER, J ;
GRUNKEMEIER, GL .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1992, 6 (07) :339-346
[5]  
HEDGER P, 1993, IMPORTANT UPDATED IN
[6]  
HEDGER P, 1992, IMPORTANT UPDATED IN
[7]  
LINDBLOM D, 1992, BRIT HEART J, V68, P249
[8]   FATIGUE LIFE ESTIMATION PROCEDURES FOR THE ENDURANCE OF A CARDIAC-VALVE PROSTHESIS - STRESS LIFE AND DAMAGE-TOLERANT ANALYSES [J].
RITCHIE, RO ;
LUBOCK, P .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1986, 108 (02) :153-160
[9]   METALLURGICAL ANALYSIS OF FAILED BJORK-SHILEY CARDIAC-VALVE PROSTHESES [J].
SACKS, SH ;
HARRISON, M ;
BISCHLER, PJE ;
MARTIN, JW ;
WATKINS, J ;
GUNNING, ALF .
THORAX, 1986, 41 (02) :142-147
[10]  
TREASURE T, 1991, BRIT HEART J, V66, P333