Electrochemical release testing of nickel-titanium orthodontic wires in artificial saliva using thin layer activation

被引:43
作者
Cioffi, M
Gilliland, D
Ceccone, G
Chiesa, R
Cigada, A
机构
[1] Politecn Milan, Dipartimento Bioingn, I-20133 Milan, Italy
[2] Commiss European Communities, Inst Hlth & Consumer Protect, I-21020 Ispra, Italy
[3] Politecn Milan, Dipartimento Chim Mat & Ingn Chim G Natta, I-20131 Milan, Italy
关键词
nickel-titanium alloy; orthodontics; corrosion; fluoride; strain;
D O I
10.1016/j.actbio.2005.07.008
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Alloys based on Ni-Ti intermetallics generally exhibit special shape memory and pseudoelastic properties, which make them desirable for use in the dental field as orthodontic wires. The possibility of nickel release from these materials is of high concern, because the allergenicity of this element. The aim of this study was to test pseudoelastic Ni-Ti wires in simulated physiological conditions, investigating the combined effect of strain and fluoridated media: the wires were examined both under strained (5% tensile strain) and unstrained conditions, in fluoridated artificial saliva at 37 degrees C. Real time electrochemical nickel release testing was performed using a novel application of a radiotracer based method, thin layer activation (TLA). TLA was validated, in unstrained conditions, against adsorptive stripping voltammetry methodology. Control tests were also performed in non-fluoridated artificial saliva. From our research it transpired that the corrosion behaviour of Ni-Ti alloy is highly affected by the fluoride content, showing a release of 4.79 +/- 0.10 mu g/cm(2)/day, but, differently from other biomaterials, it does not seem to be affected by elastic tensile strain. The application of the TLA method in the biomedical field appears a suitable technique to monitor in real time the corrosion behaviour of biomedical devices. (c) 2005 Actabio Materialia Inc. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:717 / 724
页数:8
相关论文
共 33 条
[1]
Stripping voltammetry for the determination of trace metal speciation and in-situ measurements of trace metal distributions in marine waters [J].
Achterberg, EP ;
Braungardt, C .
ANALYTICA CHIMICA ACTA, 1999, 400 :381-397
[2]
BIODEGRADATION OF ORTHODONTIC APPLIANCES .1. BIODEGRADATION OF NICKEL AND CHROMIUM INVITRO [J].
BARRETT, RD ;
BISHARA, SE ;
QUINN, JK .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1993, 103 (01) :8-14
[3]
Baumann Michael A, 2004, Dent Clin North Am, V48, P55, DOI 10.1016/j.cden.2003.11.001
[4]
Effect of surface treatment of NiTi alloy on its corrosion behavior in Hanks' solution [J].
Cisse, O ;
Savadogo, O ;
Wu, M ;
Yahia, L .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2002, 61 (03) :339-345
[5]
NITINOL INTRAVASCULAR STENT - RESULTS OF PRECLINICAL EVALUATION [J].
CRAGG, AH ;
DEJONG, SC ;
BARNHART, WH ;
LANDAS, SK ;
SMITH, TP .
RADIOLOGY, 1993, 189 (03) :775-778
[6]
Dai KR, 1996, BIO-MED MATER ENG, V6, P233
[7]
Electrochemical release testing of a stainless steel in a glucose solution using thin layer activation [J].
Garcia-Alonso, MC ;
Macchi, G ;
Brugnoni, C ;
Stroosnijder, MF .
CORROSION SCIENCE, 2002, 44 (01) :129-143
[8]
Shape memory alloys for medical applications [J].
Gil, FJ ;
Planell, JA .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1998, 212 (H6) :473-488
[9]
The use of thin layer activation in high-temperature cyclic oxidation studies [J].
Haanappel, VAC ;
Jacob, YP ;
Stroosnijder, MF .
CORROSION SCIENCE, 2002, 44 (07) :1411-1423
[10]
Hensten-Pettersen A, 1998, EUR J ORAL SCI, V106, P707