The effects of fatty deposits on the accuracy of the Fibroscan® liver transient elastography ultrasound system

被引:26
作者
Cournane, S. [1 ]
Browne, J. E. [2 ,3 ]
Fagan, A. J. [1 ,4 ]
机构
[1] St James Hosp, Dept Med Phys & Bioengn, Dublin 8, Ireland
[2] Dublin Inst Technol, Sch Phys, Med Ultrasound Phys & Technol Grp, Dublin 8, Ireland
[3] Dublin Inst Technol, FOCAS Inst, Dublin 8, Ireland
[4] St Jamess Hospital, Trinity Coll Dublin, Ctr Adv Med Imaging, Dublin, Ireland
关键词
ACOUSTIC RADIATION FORCE; CHRONIC HEPATITIS-C; NONINVASIVE ASSESSMENT; IN-VIVO; DISEASE NAFLD; FIBROSIS; PHANTOMS; FEASIBILITY; ELASTICITY; STIFFNESS;
D O I
10.1088/0031-9155/57/12/3901
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new generation of ultrasound transient elastography (TE) systems have emerged which exploit the well-known correlation between the liver's pathological and mechanical properties through measurements of the Young's elastic modulus; however, little work has been carried out to examine the effect that fatty deposits may have on the TE measurement accuracy. An investigation was carried out on the effects on the measurement accuracy of a TE ultrasound system, the Fibroscan (R), caused by overlaying fat layers of varying thickness on healthy liver-mimicking phantoms, simulating in vivo conditions for obese patients. Furthermore, a steatosis effect similar to that in non-alcoholic fatty liver disease (NAFLD) and alcoholic liver disease (ALD) was simulated to investigate its effect on the TE system. A range of novel elastography fatmimicking materials were developed using 6-10 wt% poly(vinyl alcohol) cryogel capable of achieving a range of acoustic velocities (1482-1530 m s(-1)) and attenuation coefficients (0.4-1 dB MHz(-1) cm(-1)) for simulating different liver states. Laboratory-based acoustic velocities and attenuation coefficients were measured while the Young's modulus was established through a gold standard compression testing method. A significant variation of the Young's elastic modulus was measured in healthy phantoms with overlaying fat layers of thicknesses exceeding 45 mm, impinging on the scanners region of interest, overestimating the compression tested values by up to 11 kPa in some cases. Furthermore, Fibroscan (R) measurements of the steatosis phantoms showed a consistent overestimation (similar to 54%), which strongly suggests that the speed of sound mismatch between that of liver tissue and that assumed by the scanner is responsible for the high clinical cut-offs established in the case of ALD and NAFLD.
引用
收藏
页码:3901 / 3914
页数:14
相关论文
共 41 条
[1]  
[Anonymous], ISO7743
[2]  
[Anonymous], 2014, Diagnostic Ultrasound Imaging
[3]   Assessment of liver viscoelasticity using multifrequency MR elastography [J].
Asbach, Patrick ;
Klatt, Dieter ;
Hamhaber, Uwe ;
Braun, Juergen ;
Somasundaram, Rajan ;
Hamm, Bernd ;
Sack, Ingolf .
MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (02) :373-379
[4]  
Bavu E, 2010, J HEPATOL, V52, pS59
[5]   An algorithm for the grading of activity in chronic hepatitis C [J].
Bedossa, P ;
Poynard, T .
HEPATOLOGY, 1996, 24 (02) :289-293
[6]   Investigation of the effect of subcutaneous fat on image quality performance of 2D conventional imaging and tissue harmonic imaging [J].
Browne, JE ;
Watson, AJ ;
Hoskins, PR ;
Elliott, AT .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2005, 31 (07) :957-964
[7]   Assessment of the acoustic properties of common tissue-mimicking test phantoms [J].
Browne, JE ;
Ramnarine, KV ;
Watson, AJ ;
Hoskins, PR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (07) :1053-1060
[8]   NOVEL TISSUE MIMICKING MATERIALS FOR HIGH FREQUENCY BREAST ULTRASOUND PHANTOMS [J].
Cannon, Louise M. ;
Fagan, Andrew J. ;
Browne, Jacinta E. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2011, 37 (01) :122-135
[9]   Prospective comparison of transient elastography, fibrotest, APRI, and liver biopsy for the assessment of fibrosis in chronic hepatitis C [J].
Castéra, L ;
Vergniol, J ;
Foucher, J ;
Le Bail, B ;
Chanteloup, E ;
Haaser, M ;
Darriet, M ;
Couzigou, P ;
De Lédinghen, V .
GASTROENTEROLOGY, 2005, 128 (02) :343-350
[10]   Non-invasive evaluation of liver fibrosis using transient elastography [J].
Castera, Laurent ;
Forns, Xavier ;
Alberti, Alfredo .
JOURNAL OF HEPATOLOGY, 2008, 48 (05) :835-847