FREEHAND REAL-TIME ELASTOGRAPHY: IMPACT OF SCANNING PARAMETERS ON IMAGE QUALITY AND IN VITRO INTRA- AND INTEROBSERVER VALIDATIONS

被引:117
作者
Havre, Roald Flesland [1 ,2 ]
Elde, Erlend [2 ]
Gilja, Odd Helge [1 ,2 ]
Odegaard, Svein [1 ,2 ]
Eide, Geir Egil [3 ,4 ]
Matre, Knut [2 ]
Nesje, Lars Birger [1 ,2 ]
机构
[1] Haukeland Hosp, Dept Med, Natl Ctr Ultrasound Gastroenterol, N-5021 Bergen, Norway
[2] Univ Bergen, Inst Med, Bergen, Norway
[3] Haukeland Hosp, Clin Res Ctr, N-5021 Bergen, Norway
[4] Univ Bergen, Dept Publ Hlth & Primary Hlth Care, Sect Epidemiol & Med Stat, Bergen, Norway
关键词
Real-time elastography; Ultrasonography; Free-hand application; Phantom; Strain; Elastic contrast; Elasticity dynamic range; In vitro; Interobserver agreement; Intraobserver agreement;
D O I
10.1016/j.ultrasmedbio.2008.03.009
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
Real-time elastography is a method for visualization of the elastic properties of soft tissue and may potentially enable differentiation between malignant and benign pathologic lesions. Our aim was to validate the method on a tissue-mimicking (TM) phantom and to evaluate the influence of different scanning parameters and investigator variability. A TM-phantom containing eight spherical inclusions with known storage modulus was examined using two different transducers on an ultrasound (US) scanner equipped with software for real-time elasticity imaging. The ultrasound transducers were moved vertically in a repetitive manner to induce strain. Two investigators performed series of standardized elastography scans applying a 0-4 categorical quality scale to evaluate the influence of seven parameters: dynamic range of elasticity, region-of-interest, frequency of transducer movement, rejection of elastogram noise, frame rate, persistence and smoothing. Subsequently, repeated examinations of four selected inclusions were performed using a visual analog scale (VAS) where investigators marked a 100 mm horizontal line representing the span in image quality based on experience from the first examination. The hardest and softest inclusions were imaged more clearly than the inclusions with elasticity more similar to the background material. Intraobserver agreement on elastogram quality was good (kappa: 0.67 - 0.75) and interobserver agreement average (kappa: 0.55 - 0.56) when using the categorical scale. The subsequent VAS evaluation gave intraclass-correlation coefficients for the two observers of 0.98 and 0.93, respectively, and an interclass-correlation coefficient of 0.93. Real-time elastography adequately visualized isoechoic inclusions with different elastic properties in a TM-phantom with acceptable intra- and interobserver agreement. Dynamic range of elasticity was the parameter with most impact on the elastographic visualization of inclusions. (E-mail: roald.flesland.havre@helse-bergen.no) (C) 2008 Published by Elsevier Inc. on behalf of World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1638 / 1650
页数:13
相关论文
共 47 条
[1]
ALTMAN DG, 1991, SOME COMMON PROBLEMS, P396
[2]
Measuring agreement in method comparison studies [J].
Bland, JM ;
Altman, DG .
STATISTICAL METHODS IN MEDICAL RESEARCH, 1999, 8 (02) :135-160
[3]
STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]
Assessment of the acoustic properties of common tissue-mimicking test phantoms [J].
Browne, JE ;
Ramnarine, KV ;
Watson, AJ ;
Hoskins, PR .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (07) :1053-1060
[5]
[6]
Fleiss J., 1986, Reliability of measurement: the design and analysis of clinical experiments
[7]
Reproducibility of transient elastography in the evaluation of liver fibrosis in patients with chronic liver disease [J].
Fraquelli, Mirella ;
Rigamonti, Cristina ;
Casazza, Giovanni ;
Conte, Dario ;
Donato, Maria Francesca ;
Ronchi, Guido ;
Colombo, Massimo .
GUT, 2007, 56 (07) :968-973
[8]
Frey H, 2003, RADIOLOGE, V43, P850, DOI 10.1007/s00117-003-0943-2
[9]
Elastography of breast lesions: Initial clinical results [J].
Garra, BS ;
Cespedes, EI ;
Ophir, J ;
Spratt, SR ;
Zuurbier, RA ;
Magnant, CM ;
Pennanen, MF .
RADIOLOGY, 1997, 202 (01) :79-86
[10]
Endoscopic ultrasound elastography: the first step towards virtual biopsy? Preliminary results in 49 patients [J].
Giovannini, M ;
Hookey, LC ;
Bories, E ;
Pesenti, C ;
Monges, G ;
Delpero, JR .
ENDOSCOPY, 2006, 38 (04) :344-348