Monte Carlo simulation estimates of neutron doses to critical organs of a patient undergoing 18 MV x-ray LINAC-based radiotherapy

被引:38
作者
Barquero, R [1 ]
Edwards, TM
Iñiguez, MP
Vega-Carrillo, HR
机构
[1] Univ Valladolid, Hosp Rio Hortega, Valladolid, Spain
[2] Univ Valladolid, Dept Fis Teor Atom Mol & Nucl, Valladolid, Spain
[3] UAs Estudios Nucl Ingn Elect Matemat, Zacatecas 98000, Mexico
关键词
accelerator neutron source; photon contamination; neutron dosimetry; Monte Carlo calculations; MCNP;
D O I
10.1118/1.2122547
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Absorbed photoneutron dose to patients undergoing 18 MV x-ray therapy was studied using Monte Carlo simulations based on the MCNPX code. Two separate transport simulations were conducted, one for the photoneutron contribution and another for neutron capture gamma rays. The phantom model used was of a female patient receiving a four-field pelvic box treatment. Photoneutron doses were determinate to be higher for organs and tissues located inside the treatment field, especially those closest to the patient's skin. The maximum organ equivalent dose per x-ray treatment dose achieved within each treatment port was 719 ASv/Gy to the rectum (180 degrees field), 190 ASv/Gy to the intestine wall (0 degrees field), 51 mu Sv/Gy to the colon wall (90 degrees field), and 45 mu Sv/Gy to the skin ( 270 degrees field). The maximum neutron equivalent dose per x-ray treatment dose received by organs outside the treatment field was 65 ASv/Gy to the skin in the antero-posterior field. A mean value of 5 +/- 2 mu Sv/Gy was obtained for organs distant from the treatment field. Distant organ neutron equivalent doses are all of the same order of magnitude and constitute a good estimate of deep organ neutron equivalent doses. Using the risk assessment method of the ICRP-60 report, the greatest likelihood of fatal secondary cancer for a 70 Gy dose is estimated to be 0.02% for the pelvic postero-anterior field, the rectum being the organ representing the maximum contribution of 0.011%. (c) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:3579 / 3588
页数:10
相关论文
共 16 条
[1]  
*ANS, 1999, ANSIHPSN1335
[2]   Neutron spectra and dosimetric features around an 18 MV Linac accelerator [J].
Barquero, R ;
Mendez, R ;
Vega-Carrillo, HR ;
Iñiguez, MP ;
Edwards, TM .
HEALTH PHYSICS, 2005, 88 (01) :48-58
[3]  
BARQUERO R, 2002, THESIS U VALLADOLID
[4]  
Briesmeister J.F., 2000, LA13709M LOS AL NAT
[5]   Photonuclear dose calculations for high-energy photon beams from Siemens and Varian linacs [J].
Chibani, O ;
Ma, CMC .
MEDICAL PHYSICS, 2003, 30 (08) :1990-2000
[6]   In vivo neutron dosimetry during high-energy bremsstrahlung radiotherapy [J].
D'Errico, F ;
Nath, R ;
Silvano, G ;
Tana, L .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 41 (05) :1185-1192
[7]   Contamination dose from photoneutron processes in bodily tissues during therapeutic radiation delivery [J].
Difilippo, F ;
Papiez, L ;
Moskvin, V ;
Peplow, D ;
DesRosiers, C ;
Johnson, J ;
Timmerman, R ;
Randall, M ;
Lillie, R .
MEDICAL PHYSICS, 2003, 30 (10) :2849-2854
[8]  
*ICRP, 2000, 84 ICRP
[9]  
ICRP, 1991, Report No. 60
[10]  
International Commission on Radiation Units and Measurements, 1989, 44 ICRU