Dualband spectral-spatial RF pulses for prostate MR spectroscopic imaging

被引:83
作者
Schricker, AA
Pauly, JM
Kurhanewicz, J
Swanson, MG
Vigneron, DB
机构
[1] Univ Calif San Francisco, Dept Radiol, Magnet Resonance Sci Ctr, San Francisco, CA 94143 USA
[2] Stanford Univ, Dept Elect Engn, Palo Alto, CA 94304 USA
关键词
RF pulse design; MR spectroscopic imaging; spectral localization; prostate cancer; magnetic resonance imaging;
D O I
10.1002/mrm.1302
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Although MR spectroscopic imaging (MRSI) of the prostate has demonstrated clinical utility for the staging and monitoring of cancer extent, current acquisition methods are often inadequate in several aspects. Conventional 180 degrees pulses can suffer from chemical shift misregistration, and have high peak-power requirements that can exceed hardware limits in many prostate MRSI studies. Optimal water and lipid suppression are also critical to obtain interpretable spectra. While complete suppression of the periprostatic lipid resonance is desired, controlled partial suppression of water can provide a valuable phase and frequency reference for data analysis and an assessment of experimental success in cases in which all other resonances are undetectable following treatment. In this study, new spectral-spatial RIF pulses were developed to negate chemical shift misregistration errors and to provide dualband excitation with partial excitation of the water resonance and full excitation of the metabolites of interest. Optimal phase modulation was also included in the pulse design to provide 40% reduction in peak RF power. Patient studies using the new pulses demonstrated both feasibility and clear benefits in the reliability and applicability of prostate cancer MRSI. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:1079 / 1087
页数:9
相关论文
共 28 条
[1]   SPATIAL LOCALIZATION IN NMR-SPECTROSCOPY INVIVO [J].
BOTTOMLEY, PA .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1987, 508 :333-348
[2]  
KURHANEWICZ J, 2000, PROSTATE GLAND RADIO, P115
[3]   Optimized outer volume suppression for single-shot fast spin-echo cardiac imaging [J].
Le Roux, P ;
Gilles, RJ ;
McKinnon, GC ;
Carlier, PG .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (05) :1022-1032
[4]   ACQUISITION AND QUANTITATION IN PROTON SPECTROSCOPY [J].
LUYTEN, PR ;
MARIEN, AJH ;
DENHOLLANDER, JA .
NMR IN BIOMEDICINE, 1991, 4 (02) :64-69
[5]  
Males RG, 2000, MAGNET RESON MED, V43, P17, DOI 10.1002/(SICI)1522-2594(200001)43:1<17::AID-MRM3>3.0.CO
[6]  
2-6
[7]   SIMULTANEOUS SPATIAL AND SPECTRAL SELECTIVE EXCITATION [J].
MEYER, CH ;
PAULY, JM ;
MACOVSKI, A ;
NISHIMURA, DG .
MAGNETIC RESONANCE IN MEDICINE, 1990, 15 (02) :287-304
[8]  
Moonen C T, 1989, NMR Biomed, V2, P201, DOI 10.1002/nbm.1940020506
[9]   SURFACE COIL MR-IMAGING OF THE HUMAN BRAIN WITH AN ANALYTIC RECEPTION PROFILE CORRECTION [J].
MOYHER, SE ;
VIGNERON, DB ;
NELSON, SJ .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (02) :139-144
[10]   ECHO-PLANAR SPIN-ECHO AND INVERSION PULSES [J].
PAULY, J ;
SPIELMAN, D ;
MACOVSKI, A .
MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (06) :776-782