Influence of electrode mismatch on Cole parameter estimation from Total Right Side Electrical Bioimpedance Spectroscopy measurements

被引:19
作者
Buendia, Ruben [1 ,2 ]
Bogonez-Franco, Paco [3 ]
Nescolarde, Lexa [3 ]
Seoane, Fernando [1 ,4 ]
机构
[1] Univ Boras, Sch Engn, SE-50190 Boras, Sweden
[2] Univ Alcala, Dept Signal Theory & Commun, ES-28871 Madrid, Spain
[3] Tech Univ Catalonia, Dept Elect Engn, ES-08034 Barcelona, Spain
[4] Royal Inst Technol, Sch Technol & Hlth, SE-14152 Huddinge, Sweden
关键词
Electrodes; Bioimpedance Spectroscopy; Measurement artefacts; Cole parameters; Electrode polarization impedance; BIOELECTRICAL-IMPEDANCE ANALYSIS; WHOLE-BODY; MODEL;
D O I
10.1016/j.medengphy.2012.05.011
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
Applications based on measurements of Electrical Bioimpedance (EBI) spectroscopy analysis, like assessment of body composition, have proliferated in the past years. Currently Body Composition Assessment (BCA) based in Bioimpedance Spectroscopy (BIS) analysis relays on an accurate estimation of the Cole parameters R-0 and R-infinity. A recent study by Bogonez-Franco et al. has proposed electrode mismatch as source of remarkable artefacts in BIS measurements. Using Total Right Side BIS measurements from the aforementioned study, this work has focused on the influence of electrode mismatch on the estimation of R-0 and R-infinity using the Non-Linear Least Square curve fitting technique on the modulus of the impedance. The results show that electrode mismatch on the voltage sensing electrodes produces an overestimation of the impedance spectrum leading to a wrong estimation of the parameters R-0 and R-infinity, and consequently obtaining values around 4% larger that the values obtained from BIS without electrode mismatch. The specific key factors behind electrode mismatch or its influence on the analysis of single and spectroscopy measurements have not been investigated yet, no compensation or correction technique is available to overcome the deviation produced on the EBI measurement. Since textile-enabled EBI applications using dry textrodes, i.e. textile electrodes with dry skin-electrode interfaces and potentially large values of electrode polarization impedance are more prone to produce electrode mismatch, the lack of a correction or compensation technique might hinder the proliferation of textile-enabled EBI applications for personalized healthcare monitoring. (C) 2012 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1024 / 1028
页数:5
相关论文
共 28 条
[1]
Patient-instrument connection errors in bioelectrical impedance measurement [J].
Al-Hatib, F .
PHYSIOLOGICAL MEASUREMENT, 1998, 19 (02) :285-296
[2]
Cole Equation and Parameter Estimation from Electrical Bioimpedance Spectroscopy Measurements - A Comparative Study [J].
Ayllon, David ;
Seoane, Fernando ;
Gil-Pita, Roberto .
2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, :3779-3782
[3]
Characterization of textile electrodes and conductors using standardized measurement setups [J].
Beckmann, L. ;
Neuhaus, C. ;
Medrano, G. ;
Jungbecker, N. ;
Walter, M. ;
Gries, T. ;
Leonhardt, S. .
PHYSIOLOGICAL MEASUREMENT, 2010, 31 (02) :233-247
[4]
Bogonez-Franco P, 2009, PHYSIOL MEAS, V30, P15
[5]
Sources of error in bioimpedance spectroscopy [J].
Bolton, MP ;
Ward, LC ;
Khan, A ;
Campbell, I ;
Nightingale, P ;
Dewit, O ;
Elia, M .
PHYSIOLOGICAL MEASUREMENT, 1998, 19 (02) :235-245
[6]
Experimental validation of a method for removing the capacitive leakage artifact from electrical bioimpedance spectroscopy measurements [J].
Buendia, R. ;
Seoane, F. ;
Gil-Pita, R. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2010, 21 (11)
[7]
Buendia R, 2011, J ELECT BIOIMPEDANCE, V2, P6
[8]
Hook Effect Correction & Resistance-based Cole Fitting Prior Cole Model-based Analysis: Experimental Validation [J].
Buendia, Ruben ;
Seoane, Fernando ;
Harris, Matthew ;
Caffarel, Jennifer ;
Gil, Roberto .
2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, :6563-6566
[9]
Cano JA, 2012, THESIS U BORAS BORAS
[10]
Cole K. S., 1940, Cold Spring Harbor Symposia on Quantitative Biology, V8, P110