Patient-instrument connection errors in bioelectrical impedance measurement

被引:23
作者
Al-Hatib, F [1 ]
机构
[1] Bulgarian Acad Sci, Ctr Biomed Engn, Sofia 1113, Bulgaria
关键词
bioelectrical impedance; impedance spectroscopy; instrumentation; patient-instrument interface; measurement errors;
D O I
10.1088/0967-3334/19/2/015
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Bioelectrical impedance in vivo measurement errors due to the connection between instrument and patient are analysed theoretically and experimentally. Special attention is paid to the patient cable capacitances and to the reduction of unwanted common-mode signals at the voltage amplifier input. Experiments with single-ended and symmetrical current sources, with different simulated body impedance values, with five types of cable connection and with three types of cable screen driving are carried out. The use of a balanced voltage-to-current transverter with a floating common point is recommended. Four separate patient cables should be used with separately driven screens by high-quality unity-gain buffers. Thus the impedance module measurement error can be reduced to below 1% in the range of 20 to 1000 Omega and the phase measurement error to less than 0.3 degrees, for frequencies from 1 kHz to 300 kHz.
引用
收藏
页码:285 / 296
页数:12
相关论文
共 29 条
[1]   COMPLEX BIOELECTRIC IMPEDANCE MEASUREMENT SYSTEM FOR THE FREQUENCY-RANGE FROM 5 HZ TO 1 MHZ [J].
ACKMANN, JJ .
ANNALS OF BIOMEDICAL ENGINEERING, 1993, 21 (02) :135-146
[2]  
BAKER LE, 1993, J CLIN ENG, V18, P221
[3]   Current approaches to analogue instrumentation design in electrical impedance tomography [J].
Boone, KG ;
Holder, DS .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 (04) :229-247
[4]  
BOONE KG, 1995, INNOV TECH BIOL MED, V16, P62
[5]  
BRIZZOLARA A, 1997, EUR J ENDOCRINOL, V135, P609
[6]  
Cornish BH, 1996, EUR J CLIN NUTR, V50, P159
[7]  
DEURENBERG P, 1992, EUR J CLIN NUTR, V46, P247
[8]  
Gerth Wayne A., 1993, Journal of Clinical Engineering, V18, P253
[9]   MICROPROCESSOR-BASED SYSTEM FOR MEASUREMENT OF ELECTRICAL IMPEDANCES DURING HEMODIALYSIS AND IN POSTOPERATIVE CARE [J].
GOOVAERTS, HG ;
DEVRIES, FR ;
MEIJER, JH ;
DEVRIES, PMJM ;
DONKER, AJM ;
SCHNEIDER, H .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (01) :75-80
[10]   COMPARISON OF BIOIMPEDANCE SPECTROSCOPY AND MULTIFREQUENCY BIOIMPEDANCE ANALYSIS FOR THE ASSESSMENT OF EXTRACELLULAR AND TOTAL-BODY WATER IN SURGICAL PATIENTS [J].
HANNAN, WJ ;
COWEN, SJ ;
PLESTER, CE ;
FEARON, KCH ;
DEBEAU, A .
CLINICAL SCIENCE, 1995, 89 (06) :651-658