Oxygen flux fluorescence lifetime imaging

被引:54
作者
Hartmann, P [1 ]
Ziegler, W [1 ]
Holst, G [1 ]
Lubbers, DW [1 ]
机构
[1] MAX PLANCK INST MOL PHYSIOL,D-44139 DORTMUND,GERMANY
关键词
fluorescent lifetime imaging (FLIM); oxygen; luminescent sensors; skin; phase-resolved spectroscopy;
D O I
10.1016/S0925-4005(97)80179-7
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We describe a device capable of imaging distributions of surface PO2 as well as oxygen flux (e.g., into the human skin), Oxygen is monitored by a phase-angle-sensitive imaging technique using the effect of luminescence lifetime quenching of ruthenium complexes incorporated into a polymer layer system. The dyes are excited by modulated radiation of blue light-emitting diodes. The phase-shifted luminescence is detected by an intensified CCD camera. Several modulation techniques have been comparatively investigated. The obtained images correspond pixel-by-pixel to the local luminescence lifetimes, which are a direct measure for the distribution of surface PO2, or alternatively for the oxygen flux through a permeable diffusion barrier. For an illuminated sensor area of 10 cm(2) and a spatial resolution of I mm at the sensor layer, the oxygen resolution is better than Delta PO2=0.3 torr in the absence of oxygen, and better than Delta PO2=4 torr at PO2= 160 torr.
引用
收藏
页码:110 / 115
页数:6
相关论文
共 23 条
  • [1] CLEGG RM, 1992, P SOC PHOTO-OPT INS, V1640, P448, DOI 10.1117/12.58237
  • [2] Lifetime imaging of luminescent oxygen sensors based on all-solid-state technology
    Hartmann, P
    Ziegler, W
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (24) : 4512 - 4514
  • [3] LUMINESCENCE QUENCHING BEHAVIOR OF AN OXYGEN SENSOR-BASED ON A RU(II) COMPLEX DISSOLVED IN POLYSTYRENE
    HARTMANN, P
    LEINER, MJP
    LIPPITSCH, ME
    [J]. ANALYTICAL CHEMISTRY, 1995, 67 (01) : 88 - 93
  • [4] RESPONSE CHARACTERISTICS OF LUMINESCENT OXYGEN SENSORS
    HARTMANN, P
    LEINER, MJP
    LIPPITSCH, ME
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) : 251 - 257
  • [5] Effects of polymer matrices on calibration functions of luminescent oxygen sensors based on porphyrin ketone complexes
    Hartmann, P
    Trettnak, W
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (15) : 2615 - 2620
  • [6] FLOX - AN OXYGEN-FLUX-MEASURING SYSTEM USING A PHASE-MODULATION METHOD TO EVALUATE THE OXYGEN-DEPENDENT FLUORESCENCE LIFETIME
    HOLST, GA
    KOSTER, T
    VOGES, E
    LUBBERS, DW
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1995, 29 (1-3) : 231 - 239
  • [7] Huch A, 1979, CONTINUOUS TRANSCUTA
  • [8] IN-VIVO VISUALIZATION OF OXYGEN-TRANSPORT IN MICROVASCULAR NETWORK
    ITOH, T
    YAEGASHI, K
    KOSAKA, T
    KINOSHITA, T
    MORIMOTO, T
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1994, 267 (05): : H2068 - H2078
  • [9] LUMINESCENT BAROMETRY IN WIND TUNNELS
    KAVANDI, J
    CALLIS, J
    GOUTERMAN, M
    KHALIL, G
    WRIGHT, D
    GREEN, E
    BURNS, D
    MCLACHLAN, B
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1990, 61 (11) : 3340 - 3347
  • [10] KHALIL GE, 1991, Patent No. 5043286