A new concept of integrated cardiopulmonary bypass circuit

被引:27
作者
Mueller, XM [1 ]
Jegger, D [1 ]
Augstburger, M [1 ]
Horisberger, J [1 ]
Godar, G [1 ]
von Segesser, LK [1 ]
机构
[1] CHUV, Dept Cardio Vasc Surg, CH-1011 Lausanne, Switzerland
关键词
cardiopulmonary bypass; hemodilution; gas exchange;
D O I
10.1016/S1010-7940(02)00086-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Standard cardiopulmonary bypass (CPB) circuits with their large surface area and volume contribute to postoperative systemic inflammatory reaction and hemodilution. In order to minimize these problems a new approach his been developed resulting in a single disposable, compact arterio-venous loop, which has integral kinetic-assist pumping, oxygenating, air removal, and gross filtration capabilities (CardioVention Inc., Santa Clara, CA, USA). The impact of this system on gas exchange capacity, blood elements and hemolysis is compared to that of a conventional circuit in a model of prolonged perfusion. Methods: Twelve calves (mean body weight: 72.2 +/- 3.7 kg) were placed on cardiopulmonary bypass for 6 It with a flow of 51/min, and randomly assigned to the CardioVention system (n = 6) or a standard CPB circuit (n = 6). A standard battery of blood samples was taken before bypass and throughout bypass. Analysis of variance was used for comparison. Results: The hematocrit remained stable throughout the experiment in the CardioVention group, whereas it dropped in the standard group in the early phase of perfusion. When normalized for prebypass values, both profiles differed significantly (P < 0.01). Both O-2 and CO2 transfers were significantly improved in the CardioVention group (P = 0.04 and P < 0.001, respectively). There was a slightly higher pressure drop in the CardioVention group but no single value exceeded 112 mmHg. No hemolysis could be detected in either group with all free plasma Hb values below 15 mg/l. Thrombocyte count, when corrected by hematocrit and normalized by prebypass values, exhibited an increased drop in the standard group (P = 0.03). Conclusion: The CardioVention system with its concept of limited priming volume and exposed foreign surface area, improves gas exchange probably because of the absence of detectable hemodilution, and appears to limit the decrease in the thrombocyte count which may be ascribed to the reduced surface. Despite the volume and surface constraints, no hemolysis could be detected throughout the 6 h full-flow perfusion period. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:840 / 846
页数:7
相关论文
共 16 条
[1]  
ADDONIZIO VP, 1979, J THORAC CARDIOV SUR, V77, P119
[2]   Lung injury and acute respiratory distress syndrome after cardiopulmonary bypass [J].
Asimakopoulos, G ;
Smith, PLC ;
Ratnatunga, CP ;
Taylor, KM .
ANNALS OF THORACIC SURGERY, 1999, 68 (03) :1107-1115
[3]   BRAIN EMBOLIC PHENOMENA ASSOCIATED WITH CARDIOPULMONARY BYPASS [J].
CHALLA, VR ;
MOODY, DM ;
TROOST, BT .
JOURNAL OF THE NEUROLOGICAL SCIENCES, 1993, 117 (1-2) :224-231
[4]   Size makes a difference: use of a low-prime cardiopulmonary bypass circuit and autologous priming in small adults [J].
Cormack, JE ;
Forest, RJ ;
Groom, RC ;
Morton, J .
PERFUSION-UK, 2000, 15 (02) :129-135
[5]  
Fang WC, 1997, CIRCULATION, V96, P194
[6]   EFFECTS OF SILICONE ANTIFOAM ON PULMONARY VASCULATURE [J].
GUPTA, S ;
MAITRA, TK ;
DANDAPAT, R .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1972, 63 (05) :714-&
[7]  
MILLER BE, 1997, J CARDIOTHOR VASC AN, V11, P357
[8]  
Mueller X M, 2000, J Extra Corpor Technol, V32, P152
[9]   Smart suction device for less blood trauma: a comparison with Cell Saver [J].
Mueller, XH ;
Tevaearai, HT ;
Horisberger, J ;
Augstburger, M ;
Boone, Y ;
von Segesser, LK .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 19 (04) :507-511
[10]   Hemolysis and hematology profile during perfusion: inter-species comparison [J].
Mueller, XM ;
Tevaearai, HT ;
Jegger, D ;
Tucker, O ;
von Segesser, LK .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2001, 24 (02) :89-94