Effect of RF coil excitation on field inhomogeneity at ultra high fields: A field optimized TEM resonator

被引:176
作者
Ibrahim, TS
Lee, R
Baertlein, BA
Abduljalil, AM
Zhu, H
Robitaille, PML [1 ]
机构
[1] Ohio State Univ, Dept Radiol, Columbus, OH 43210 USA
[2] Ohio State Univ, Dept Elect Engn, Columbus, OH 43210 USA
关键词
TEM resonator; B-1; field; numerical optimization; ultra high field;
D O I
10.1016/S0730-725X(01)00404-0
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this work, computational methods were utilized to optimize the field produced by the transverse electromagnetic (TEM) resonator in the presence of the human head at 8 Tesla. Optimization was achieved through the use of the classical finite difference time domain (FDTD) method and a TEM resonator loaded with an anatomically detailed human head model with a resolution of 2 mm X 2 mm X 2 mm. The head model was developed from 3D MR images. To account for the electromagnetic interactions between the coil and the tissue, the coil and the head were treated as a single system at all the steps of the model including, numerical tuning and excitation. In addition to 2, 3, 4, 6, and 10-port excitations, an antenna array concept was utilized by driving all the possible ports (24) of a 24-strut TEM resonator. The results show that significant improvement in the circularly polarized component of the transverse magnetic field could be obtained when using multiple ports and variable phase and fixed magnitude, or variable phase and variable magnitude excitations. (C) 2001 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:1339 / 1347
页数:9
相关论文
共 25 条
[1]  
[Anonymous], 1997, Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies
[2]   Theoretical model for an MRI radio frequency resonator [J].
Baertlein, BA ;
Özbay, Ö ;
Ibrahim, TS ;
Lee, R ;
Yu, Y ;
Kangarlu, A ;
Robitaille, PML .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2000, 47 (04) :535-546
[3]   A PERFECTLY MATCHED LAYER FOR THE ABSORPTION OF ELECTROMAGNETIC-WAVES [J].
BERENGER, JP .
JOURNAL OF COMPUTATIONAL PHYSICS, 1994, 114 (02) :185-200
[4]  
Bomsdorf H, 1988, NMR Biomed, V1, P151, DOI 10.1002/nbm.1940010308
[5]  
BRIDGES JF, 1988, Patent No. 4751464
[6]   NUMERICAL-SIMULATION OF ANNULAR-PHASED ARRAYS OF DIPOLES FOR HYPERTHERMIA OF DEEP-SEATED TUMORS [J].
CHEN, JY ;
GANDHI, OP .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (03) :209-216
[7]   NUMERICAL OPTIMIZATION OF 3-D SAR DISTRIBUTIONS IN CYLINDRICAL MODELS FOR ELECTROMAGNETIC HYPERTHERMIA [J].
CHOWDHURY, DQ ;
HILL, SC .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (12) :1246-1255
[8]   OPTIMIZED SMALL-BORE, HIGH-PASS RESONATOR DESIGNS [J].
CROZIER, S ;
LUESCHER, K ;
FORBES, LK ;
DODDRELL, DM .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1995, 109 (01) :1-11
[9]   An analysis and optimization of elliptical RF probes used in magnetic resonance imaging [J].
Forbes, LK ;
Crozier, S ;
Doddrell, DM .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1996, 7 (09) :1281-1290
[10]   AN EFFICIENT, HIGHLY HOMOGENEOUS RADIOFREQUENCY COIL FOR WHOLE-BODY NMR IMAGING AT 1.5-T [J].
HAYES, CE ;
EDELSTEIN, WA ;
SCHENCK, JF ;
MUELLER, OM ;
EASH, M .
JOURNAL OF MAGNETIC RESONANCE, 1985, 63 (03) :622-628