Objectives: We have explored the usefulness of an on-table, cross-sectional radiological imaging (DynaCT) in endovascular aortic repair (EVAR). DynaCT images were compared to images from a regular multidetector (16 slice) CT. In the comparison, we tested the accordance of firstly 5 relevant clinical measurements and secondly the visibility of 9 anatomical areas in the two different types of images. This imaging was carried out in addition to the usual angiographic imaging. Design, material and method: 20 patients with infrarenal abdominal aortic aneurysm (AAA) were prospectively enrolled in the study. We compared Images from DynaCT with two different doses of contrast medium to MDCT-images in two different ways. Firstly relevant arterial diameters and lengths and secondly, 9 anatomical areas were evaluated regarding visibility which was scored on a 4-point scale. Results: There were no significant differences in the measured arterial diameters and lengths. MDCT had a significantly higher visibility score than both DynaCT investigations. However, with the highest contrast medium dose we found acceptable diagnostic quality in 78-94% of the cases for 8 of the 9 investigated anatomical areas. Conclusion: Our findings indicate that on-table DynaCT are of sufficient quality to give relevant information of arterial measurements, needed in endovascular repair of infrarenal aortic aneurysms. (c) 2008 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:23 / 30
页数:8
相关论文
共 24 条
[1]   Aneurysmal sizing after endovascular repair in patients with abdominal aortic aneurysm: interobserver variability of various measurement protocols and its clinical relevance [J].
Abada, HT ;
Sapoval, MR ;
Paul, JF ;
de Maertelaer, V ;
Mousseaux, E ;
Gaux, JC .
EUROPEAN RADIOLOGY, 2003, 13 (12) :2699-2704
[2]   Emergency endovascular repair for ruptured abdominal aortic aneurysms: Feasibility and comparison of early results with conventional open repair [J].
Alsac, JM ;
Desgranges, P ;
Kobeiter, H ;
Becquemin, JP .
EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2005, 30 (06) :632-639
[3]   Infrarenal abdominal aortic aneurysm repair: Time-trends during a 20-year period [J].
Berge, Camilla ;
Haug, Erik S. ;
Romundstad, Pal R. ;
Lange, Conrad ;
Myhre, Hans O. .
WORLD JOURNAL OF SURGERY, 2007, 31 (08) :1682-1686
[4]  
BIASI L, 2008, CARDIOVASC INTE 0726
[5]   Two-year outcomes after conventional or endovascular repair of abdominal aortic aneurysms [J].
Blankensteijn, JD ;
de Jong, SECA ;
Prinssen, M ;
van der Ham, AC ;
Buth, J ;
van Sterkenburg, SMM ;
Verhagen, HJM ;
Buskens, E ;
Grobbee, DE .
NEW ENGLAND JOURNAL OF MEDICINE, 2005, 352 (23) :2398-2405
[6]  
Bongartz G., 2000, EUROPEAN GUIDELINES
[7]   Angiographic computed tomography is comparable to multislice computed tomography in lumbar myelographic imaging [J].
Buhk, Jan-Hendrik ;
Elof, Erck ;
Knauth, Michael .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2006, 30 (05) :739-741
[8]  
Cotroneo AR, 2006, RADIOL MED, V111, P597, DOI 10.1007/s11547-006-0054-z
[9]   Initial observations of endovascular aneurysm repair using Dyna-CT [J].
Eide, Kari Ravn ;
Odegard, Asbjorn ;
Myhre, Hans Olav ;
Haraldseth, Olav .
JOURNAL OF ENDOVASCULAR THERAPY, 2007, 14 (01) :50-53
[10]   Endovascular aneurysm repair versus open repair in patients with abdominal aortic aneurysm (EVAR trial 1): randomised controlled trial [J].
Greenhalgh, RM ;
Brown, LC ;
Epstein, D ;
Kwong, GPS ;
Powell, JT ;
Sculpher, MJ ;
Thompson, SG .
LANCET, 2005, 365 (9478) :2179-2186