Bedside waveforms interpretation as a tool to identify patient-ventilator asynchronies

被引:146
作者
Georgopoulos, D [1 ]
Prinianakis, G [1 ]
Kondili, E [1 ]
机构
[1] Univ Crete, Intens Care Med Dept, Univ Hosp Heraklion, Iraklion, Greece
关键词
assisted mechanical ventilation; bedside waveform interpretation; flow; volume; airway signals; patient-ventilator interaction; respiratory system mechanics; ventilator triggering;
D O I
10.1007/s00134-005-2828-5
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: During assisted modes of ventilatory support the ventilatory output is the final expression of the interaction between the ventilator and the patient's controller of breathing. This interaction may lead to patient-ventilator asynchrony, preventing the ventilator from achieving its goals, and may cause patient harm. Flow, volume, and airway pressure signals are significantly affected by patient-ventilator interaction and may serve as a tool to guide the physician to take the appropriate action to improve the synchrony between patient and ventilator. This review discusses the basic waveforms during assisted mechanical ventilation and how their interpretation may influence the management of ventilated patients. The discussion is limited on waveform eye interpretation of the signals without using any intervention which may interrupt the process of mechanical ventilation. Discussion: Flow, volume, and airway pressure may be used to (a) identify the mode of ventilator assistance, triggering delay, ineffective efforts, and autotriggering, (b) estimate qualitatively patient's respiratory efforts, and (c) recognize delayed and premature opening of exhalation valve. These signals may also serve as a tool for gross estimation of respiratory system mechanics and monitor the effects of disease progression and various therapeutic interventions. Conclusion: Flow, volume, and airway pressure waveforms are valuable real-time tools in identifying various aspects of patient-ventilator interaction.
引用
收藏
页码:34 / 47
页数:14
相关论文
共 67 条
[1]  
Aerts JGJV, 1997, EUR RESPIR J, V10, P550
[2]   FLOW AND VOLUME DEPENDENCE OF RESPIRATORY SYSTEM MECHANICS DURING CONSTANT FLOW VENTILATION IN NORMAL SUBJECTS AND IN ADULT RESPIRATORY-DISTRESS SYNDROME [J].
AULER, JOC ;
SALDIVA, PHN ;
MARTINS, MA ;
CARVALHO, CRR ;
NEGRI, EM ;
HOELZ, C ;
ZIN, WA .
CRITICAL CARE MEDICINE, 1990, 18 (10) :1080-1086
[3]   LUNG AND CHEST-WALL MECHANICAL-PROPERTIES BEFORE AND AFTER CARDIAC-SURGERY WITH CARDIOPULMONARY BYPASS [J].
BARNAS, GM ;
WATSON, RJ ;
GREEN, MD ;
SEQUEIRA, AJ ;
GILBERT, TB ;
KENT, J ;
VILLAMATER, E .
JOURNAL OF APPLIED PHYSIOLOGY, 1994, 76 (01) :166-175
[4]  
BAUM M, 1989, ANAESTHESIST, V38, P452
[5]   Increased initial flow rate reduces inspiratory work of breathing during pressure support ventilation in patients with exacerbation of chronic obstructive pulmonary disease [J].
Bonmarchand, G ;
Chevron, V ;
Chopin, C ;
Jusserand, D ;
Girault, C ;
Moritz, F ;
Leroy, J ;
Pasquis, P .
INTENSIVE CARE MEDICINE, 1996, 22 (11) :1147-1154
[6]   COMPARISON OF PRESSURE AND FLOW TRIGGERING SYSTEMS DURING CONTINUOUS POSITIVE AIRWAY PRESSURE [J].
BRANSON, RD ;
CAMPBELL, RS ;
DAVIS, K ;
JOHNSON, DJ .
CHEST, 1994, 106 (02) :540-544
[7]  
Branson Richard, 2004, Curr Opin Crit Care, V10, P23, DOI 10.1097/00075198-200402000-00005
[8]  
Branson Richard D, 2005, Respir Care Clin N Am, V11, P119, DOI 10.1016/j.rcc.2005.02.002
[9]  
BROCHARD L, 1995, PRINCIPLES PRACTICE, P239
[10]   Patient-ventilator trigger asynchrony in prolonged mechanical ventilation [J].
Chao, DC ;
Scheinhorn, DJ ;
StearnHassenpflug, M .
CHEST, 1997, 112 (06) :1592-1599