Evaluation of a desolvating microconcentric nebulizer in inductively coupled plasma mass spectrometry to improve the determination of arsenic in steels

被引:13
作者
Coedo, AG [1 ]
Dorado, T [1 ]
Padilla, I [1 ]
机构
[1] CSIC, Ctr Nacl Invest Met, Madrid 28040, Spain
关键词
desolvating microconcentric nebulizer; inductively coupled plasma mass spectrometry; steels; arsenic;
D O I
10.1366/0003702991947630
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 [仪器科学与技术]; 080401 [精密仪器及机械]; 081102 [检测技术与自动化装置];
摘要
A desolvating microconcentric nebulizer (D-MCN) was used in inductively coupled plasma mass spectrometry (ICP-MS) in order to improve the determination of As in steels. To dissolve the samples, we applied a hydrochloric-nitrit-hydrofluoric acid decomposition procedure. The isobaric interference of (ArCl+)-Ar-40-Cl-35 due to the presence of Cl from the hydrochloric acid was minimized by using the desolvating device. Compared to results achieved with cross-flow nebulization, the As sensitivity was found to be 2.4 times higher, while the depression matrix effect caused by 0.1% m/v of Pc in the As signal was 3 times lower. The limit of quantification (LOQ) calculated as the concentration corresponding to 10 times the standard deviation of a procedural blank solution (0.1% m/v pure iron solution) was 0.12 mu g g(-1) (10 times better than that for the cross-flow nebulizer), and the background equivalent concentration (BEC) was about 25 times lower due to the absence of the most significant polyatomic interferences. The relative standard deviation calculated from six replicates at a concentration level of 10 times the detection limit was 0.8%. Three certified reference materials from the National Institute of Standard and Technology-NIST 365, 2165, and 2167-were analyzed; and recovery tests were performed on pure iron solutions spiked with different As contents. The close agreement of the values demonstrates the accuracy of the method.
引用
收藏
页码:974 / 978
页数:5
相关论文
共 20 条
[1]
Aerosol-phase assisted sample digestion for the determination of trace metals in organic samples by plasma atomic emission spectrometry [J].
Allen, LB ;
Siitonen, PH ;
Thompson, HC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (07) :529-532
[2]
Determination of rare earth elements in wine by inductively coupled plasma mass spectrometry using a microconcentric nebulizer [J].
Augagneur, S ;
Medina, B ;
Szpunar, J ;
Lobinski, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) :713-721
[3]
BETTINELLI M, 1994, ATOM SPECTROSC, V15, P115
[4]
Chen HY, 1997, ATOM SPECTROSC, V18, P29
[5]
COEDO AG, 1998, 18207 EUR
[6]
DEBRAH E, 1995, ATOM SPECTROSC, V16, P197
[7]
Toward the next generation of atomic mass spectrometers - Plenary lecture [J].
Hieftje, GM ;
Myers, DP ;
Li, GQ ;
Mahoney, PP ;
Burgoyne, TW ;
Ray, SJ ;
Guzowski, JP .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :287-292
[8]
HILL SJ, 1996, J ANAL ATOM SPECTROM, V11, pR309
[9]
*INT STAND ORG, 1988, PROC ACT ISOTC 17SC1
[10]
*INT STAND ORG, 1998, STEEL IR DET ARS CON