Novel and diagnostically applicable information from optical waveform analysis of blood coagulation in disseminated intravascular coagulation

被引:55
作者
Downey, C [1 ]
Kazmi, R [1 ]
Toh, CH [1 ]
机构
[1] ROYAL LIVERPOOL UNIV HOSP,DEPT HAEMATOL,LIVERPOOL L7 8XP,MERSEYSIDE,ENGLAND
关键词
disseminated intravascular coagulation; optical profile; coagulation assays; automation;
D O I
10.1046/j.1365-2141.1997.1062972.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Transmittance waveform is the term applied to the optical profile generated from the process of clot formation on standard coagulation tests run on the MDA-180, a new-generation automated coagulation analyser. In patients with disseminated intravascular coagulation, a characteristically abnormal 'biphasic change' is seen on both the activated partial thromboplastin time and thrombin time waveforms. Increasing sleepness of the initial slope on the waveforms correlates with clinical deterioration and fulminant progression. Although the mechanism underlying the biphasic appearance remains to be elucidated, its identification provides the diagnostic laboratory with a simple, rapid and robust assay for disseminated intravascular coagulation that can help the clinician with urgent and appropriate therapeutic interventions.
引用
收藏
页码:68 / 73
页数:6
相关论文
共 8 条
[1]  
BAUER KA, 1987, BLOOD, V70, P343
[2]   DISSEMINATED INTRA-VASCULAR COAGULATION - REAPPRAISAL [J].
COLMAN, RW ;
ROBBOY, SJ ;
MINNA, JD .
ANNUAL REVIEW OF MEDICINE, 1979, 30 :359-374
[3]  
FEINSTEIN DI, 1982, BLOOD, V60, P284
[4]   Predicting the presence of plasma heparin using neural networks to analyze coagulation screening assay optical profiles [J].
Givens, TB ;
Braun, P ;
Fischer, TJ .
COMPUTERS IN BIOLOGY AND MEDICINE, 1996, 26 (06) :463-476
[5]  
HANADA T, 1985, AM J PEDIAT HEMATOL, V7, P3
[6]  
HEEB MJ, 1989, BLOOD, V73, P455
[7]  
SPERO JA, 1980, THROMB HAEMOSTASIS, V43, P28
[8]  
TEITEL JM, 1982, BLOOD, V59, P1086