A GC/MS/MS screening method for multiple organic acidemias from urine specimens

被引:43
作者
Hagen, T
Korson, MS
Sakamoto, M
Evans, JE
机构
[1] Childrens Hosp, Dept Lab Med, Boston, MA 02115 USA
[2] Childrens Hosp, Div Genet, Boston, MA 02115 USA
[3] Eunice Kennedy Shriver Ctr Mental Retardat Inc, Waltham, MA 02452 USA
关键词
gas chromatography tandem mass spectrometry; ion trap GC MS; organic acidemias; screening;
D O I
10.1016/S0009-8981(99)00037-6
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A gas chromatography tandem mass spectrometry method using an ion trap GC/MS system was developed to quickly screen urine samples for 14 organic acids associated with multiple organic acidemias. The following organic acids are used as diagnostic markers: methylmalonic acid, glutaric acid, 2-ketoisocaproic acid, succinylacetone, 3-methylcrotonylglycine, tiglylglycine, isovalerylglycine, fumaric acid, butyrylglycine, propionylglycine, hexanoylglycine, adipic acid, suberic acid, and sebacic acid. 2-ketocaproic acid is used as an internal standard. The samples are prepared using a solid-phase extraction and converted to trimethylsilyl derivatives. The extraction efficiency for the 14 compounds is between 57 and 106%. A derivatized standard mixture of the 14 markers is run prior to the patient samples to determine the accurate absolute and relative retention times. The samples are then injected and the product ion spectra monitored. For data analysis, one characteristic product ion plot is extracted for each, of the 14 marker compounds, and the presence of a peak with the expected retention time is determined. The areas of the product ion peaks are compared with the reference range determined from 30 normal controls. Ten samples of patients with known organic acidemias were measured. For all patients, diagnostic peaks at the expected retention times of at least five times the upper limit of the reference range were detected. The method, with its relatively fast sample preparation, short 10.0 min run time and simple data analysis, is suitable for use as a quick metabolic screen of very sick:patients in whom there is concern regarding the possibility of a treatable inborn error. (C) 1999 Elsevier Science BN. All rights reserved.
引用
收藏
页码:77 / 88
页数:12
相关论文
共 8 条
[1]  
Herbst RM, 1939, ORG SYNTH, V19, P4
[2]   GAS-CHROMATOGRAPHIC STUDIES OF 20 METABOLICALLY IMPORTANT ACYLGLYCINES [J].
RAMSDELL, HS ;
TANAKA, K .
CLINICA CHIMICA ACTA, 1977, 74 (02) :109-114
[3]   Determination of serum physiological concentration of methylmalonic acid by gas chromatography mass spectrometry with selected ion monitoring [J].
Rifai, N ;
Hagen, T ;
Bradley, L ;
Sakamoto, M .
ANNALS OF CLINICAL BIOCHEMISTRY, 1998, 35 :633-636
[4]   ANALYSIS OF PESTICIDE-RESIDUES IN FOOD USING GAS-CHROMATOGRAPHY TANDEM MASS-SPECTROMETRY WITH A BENCHTOP ION-TRAP MASS-SPECTROMETER [J].
SCHACHTERLE, S ;
BRITTAIN, RD ;
MILLS, JD .
JOURNAL OF CHROMATOGRAPHY A, 1994, 683 (01) :185-193
[5]   Pesticide residue analysis in fresh produce by gas chromatography - Tandem mass spectrometry [J].
Schachterle, S ;
Feigel, C .
JOURNAL OF CHROMATOGRAPHY A, 1996, 754 (1-2) :411-422
[6]   SCREENING NEWBORNS FOR MULTIPLE ORGANIC ACIDURIAS IN DRIED FILTER-PAPER URINE SAMPLES - METHOD DEVELOPMENT [J].
TUCHMAN, M ;
MCCANN, MT ;
JOHNSON, PE ;
LEMIEUX, B .
PEDIATRIC RESEARCH, 1991, 30 (04) :315-321
[7]   Evaluation of a benchtop ion trap gas chromatographic tandem mass spectrometric instrument for the analysis of a model drug, tebufelone, in plasma using a stable-isotope internal standard [J].
Wehmeyer, KR ;
Knight, PM ;
Parry, RC .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 676 (01) :53-59
[8]   FIELD-MODULATED SELECTIVE ION STORAGE IN A QUADRUPOLE ION-TRAP [J].
WELLS, G ;
HUSTON, C .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1995, 6 (10) :928-935