A rational approach to dose reduction in CT: individualized scan protocols

被引:82
作者
Wilting, JE
Zwartkruis, A
van Leeuwen, MS
Timmer, J
Kamphuis, AGA
Feldberg, M
机构
[1] Philips Med Syst Nederland BV, NL-5680 DA Best, Netherlands
[2] Univ Utrecht, Med Ctr, Utrecht, Netherlands
[3] Rode Kruis Ziekenhuis, Beverwijk, Netherlands
关键词
abdomen; spiral CT; dose reduction; image quality; technical aspects;
D O I
10.1007/s003300101039
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to demonstrate that dose reduction and constant image quality can be achieved by adjusting X-ray dose to patient size. To establish the relation between patient size, image quality and dose we scanned 19 patients with reduced dose. Image noise was measured. Four radiologists scored image quality subjectively, whereby a higher score meant less image quality. A reference patient diameter was determined for which the dose was just sufficient. Then 22 patients were scanned with the X-ray dose adjusted to their size. Again, image noise was measured and subjective image quality was scored. The dose reduction compared with the standard protocol was calculated. In the first group the measured noise was correlated to the patient diameter (rho = 0.78). This correlation is lost in the second group (rho = -0.13). The correlation between patient diameter and subjective image quality scores changes from rho = 0.60 (group 1) to rho = -0.69 (group 2). Compared with the standard protocol, the dose was reduced (mean 28 %, range 0-76 %) in 19 of : 2 patients (86 %). Dose reduction and constant noise can be achieved when the X-ray dose is adjusted to the patient diameter. With constant image noise the subjective image quality increases with larger patients.
引用
收藏
页码:2627 / 2632
页数:6
相关论文
共 15 条
[1]  
EINDHOVEN BV, 1996, TOMOSCAN AV OPERATOR, P10
[2]  
*EUR COMM STUD GRO, 1999, 16262 EUR EN
[3]  
*GE MED SYST, 1994, CT HISPEED ADVANTAGE, V3, P15
[4]  
Giacomuzzi S M, 1996, Aktuelle Radiol, V6, P110
[5]   Dose reduction in computed tomography by attenuation-based on-line modulation of tube current: evaluation of six anatomical regions [J].
Greess, H ;
Wolf, H ;
Baum, U ;
Lell, M ;
Pirkl, M ;
Kalender, W ;
Bautz, WA .
EUROPEAN RADIOLOGY, 2000, 10 (02) :391-394
[6]   THE EFFECT OF MAS VARIATION UPON COMPUTED-TOMOGRAPHY IMAGE QUALITY AS EVALUATED BY INVIVO AND INVITRO STUDIES [J].
HAAGA, JR ;
MIRALDI, F ;
MACINTYRE, W ;
LIPUMA, JP ;
BRYAN, PJ ;
WIESEN, E .
RADIOLOGY, 1981, 138 (02) :449-454
[7]   Image quality and dose in computed tomography [J].
Jurik, AG ;
Jessen, KA ;
Hansen, J .
EUROPEAN RADIOLOGY, 1997, 7 (01) :77-81
[8]   Dose reduction in CT by on-line tube current control: principles and validation on phantoms and cadavers [J].
Kalender, WA ;
Wolf, H ;
Suess, C ;
Gies, M ;
Greess, H ;
Bautz, WA .
EUROPEAN RADIOLOGY, 1999, 9 (02) :323-328
[9]   RADIATION-DOSE REDUCTION IN CT OF THE PEDIATRIC PELVIS [J].
KAMEL, IR ;
HERNANDEZ, RJ ;
MARTIN, JE ;
SCHLESINGER, AE ;
NIKLASON, LT ;
GUIRE, KE .
RADIOLOGY, 1994, 190 (03) :683-687
[10]   Simulated dose reduction in conventional chest CT: Validation study [J].
Mayo, JR ;
Whittall, KP ;
Leung, AN ;
Hartman, TE ;
Park, CS ;
Primack, SL ;
Chambers, GK ;
Limkeman, MK ;
Toth, TL ;
Fox, SH .
RADIOLOGY, 1997, 202 (02) :453-457